AD-4408C Weighing Indicator
Page
58
11. Setting List
When performing maintenance, use the following list as a memo.
In case of inquiry about the product, inform of the user setting.
11.1. Basic Functions
Function No.
Parameter or
Setting range
Description
Default
value
User
setting
F
fncf01
000000
to
111111
Disable key switch
0
: Does not disable the key switch function.
1
: Disables the key switch function.
000000
(binary)
fncf02
0
to
6
Assigns a function to the F key.
0
: None
1
: Manual print command
2
: Hold
3
: Alternate switch
4
: Momentary switch
5
: Clear the tare weight
6
: Clear the zero value
0
fncf03
1
to
3
Display update rate
1
: 20 times/second
2
: 10 times/second
3
: 5 times/second
1
fncf04
0
to
3
Assigns a function to the
display.
0
: None
1
: Zero tracking in progress
2
: Alarm (Zero range setting error, over,
low battery)
3
: F key status
0
fncf05
0
to
9
Digital filter (cutoff frequency)
0
: None
1
: 11.0 Hz
2
: 8.0 Hz
3
: 5.6 Hz
4
: 4.0 Hz
5
: 2.8 Hz
6
: 2.0 Hz
7
: 1.4 Hz
8
: 1.0 Hz
9
: 0.7 Hz
8
fncf06
1
to
10
Frequency divider ratio
1
fncf07
1
to
2
Hold
1
: Normal hold
2
: Peak hold
1
Summary of Contents for AD-4408C
Page 1: ...AD 4408C Weighing Indicator INSTRUCTION MANUAL 1WMPD4001849A ...
Page 64: ...AD 4408C Weighing Indicator Page 62 12 External Dimensions Unit mm Fig 16 External dimensions ...
Page 65: ...Page 63 AD 4408C Weighing Indicator MEMO ...
Page 66: ...AD 4408C Weighing Indicator Page 64 MEMO ...
Page 67: ......