Step 8 Press the ESC key.
Cal5et
is displayed and the calibration data is written
into the EEPROM.
The current state is the same as that of step 2. To return to the weighing mode,
press the ESC key again.
NOTE: If
C err
X is displayed, an error has occurred. Refer to “7.4. Calibration
Errors” to take some measures.
If the decimal point blinks, it indicates that the current value is not the
weight value.