![A&D AD-4408C Instruction Manual Download Page 35](http://html1.mh-extra.com/html/aandd/ad-4408c/ad-4408c_instruction-manual_446087035.webp)
Page
33
AD-4408C Weighing Indicator
8.2. Basic Functions (
fncf
)
Function No.
Parameter or
Setting range
Function Description Default
value
F
fncf01
000000
to
111111
Disable
key switch
Each digit of the setting corresponds to a key
switch. Only available in the weighing mode.
0
: Does not disable the key switch function.
1
: Disables the key switch function.
Key assignment
6
th
5
th
4
th
3
rd
2
nd
1
st
000000
(binary)
fncf02
0
to
6
F key
Assigns a function to the F key.
0
: None
1
: Manual print command
2
: Hold
3
: Alternate switch
4
: Momentary switch
5
: Clear the tare weight
6
: Clear the zero value
NOTE: Clearing the zero value can be
enabled or disabled by
Calf15
0
fncf03
1
to
3
Display
update
rate
1
: 20 times/second
2
: 10 times/second
3
: 5 times/second
1
fncf04
0
to
3
display
Assigns a function to the
display.
0
: None
1
: Zero tracking in progress
2
: Alarm (Zero range setting error, over,
low battery)
3
: F key status
0
fncf05
0
to
9
Digital
filter
Selects a cutoff frequency.
0
: None
1
: 11.0 Hz
2
: 8.0 Hz
3
: 5.6 Hz
4
: 4.0 Hz
5
: 2.8 Hz
6
: 2.0 Hz
7
: 1.4 Hz
8
: 1.0 Hz
9
: 0.7 Hz
8
fncf06
1
to
10
Frequency
divider
ratio
Divides digital filter cutoff frequency by the
setting.
1
fncf07
1
to
2
Hold
1
: Normal hold
2
: Peak hold
1
ZERO
TARE
Net/B/G
NET GROSS
ON/OFF
ESC
ENTER
Summary of Contents for AD-4408C
Page 1: ...AD 4408C Weighing Indicator INSTRUCTION MANUAL 1WMPD4001849A ...
Page 64: ...AD 4408C Weighing Indicator Page 62 12 External Dimensions Unit mm Fig 16 External dimensions ...
Page 65: ...Page 63 AD 4408C Weighing Indicator MEMO ...
Page 66: ...AD 4408C Weighing Indicator Page 64 MEMO ...
Page 67: ......