15.
Relay Output & Control Input (OP-05)
Replacing the RS-232C interface with this option, 3-relay outputs and 3-control inputs
can be used with the RS-232C interface of this option.
RS-232C functions and pin connections are the same as the RS-232C interface
described in “13. RS-232C Interface (OP-04, OP-05, OP-08)”.
The control inputs can control the indicator from an external terminal just like the front
panel key operations.
Set the external control functions at F13 - F15 of the F-Functions.
When connecting each function pin to the common pin, the indicator makes the action.
Keep a signal width of more than 100 ms for the On-time and Off-time.
ON
(Make)
More than 100 ms More than 100 ms
OFF
(Break)
The relays output the result of comparison.
Solid-state-relay
Maximum voltage
50VDC
Maximum current
100 mA DC
Maximum resistance 8
Pin connections (DIN 8 pin connector)
Function Pin No. Signal name Direction Description
1
HI
Output
Relay output HI
6
OK
Output
Relay output OK
4
LO
Output
Relay output LO
Relay
output
8
COM (out)
-
Relay common terminal
3
EXT1
Input
Control input 1 (F13)
5
EXT2
Input
Control input 2 (F14)
7
EXT3
Input
Control input 3 (F15)
Control
input
2
COM (in)
-
Control input common
Adaptable connector
JA-TCP0586 (provided with the OP-05)
Circuit
HI
OK
LO
COM (out
)
i
EXT3
EXT2
EXT1
+7
~
9V
680
Ω
680
Ω
680
Ω
COM
(in)
DIN connector
of OP-05
Control input
Relay output
15. Relay Output & Control Input (OP-05)