Weighing a Mass less than the Maximum Capacity
Step12 Set a mass value using the following keys.
key .... To select a digit.
key .... To select a value.
Step13 Place a mass equivalent to the displayed value on
the weighing unit. Proceed to step 15.
Weighing a Maximum Capacity Mass
Step14 Place a mass equivalent to the maximum capacity on
the weighing unit.
Step15 Press the ENTER key after the STABLE mark
turns on. Proceed to step 17.
ESC key ......... To proceed to step 17 without
changing the span parameter.
Caution Do not press the
ENTER
key while the STABLE
mark is off (detecting motion). Arrange the
condition using the F00 filter function.
Digital Input
Step16 Pressing the
+/-
key, a stored input voltage
parameter of the span is displayed in the unit
of mV/V. Adjust the input voltage using the
following keys.
key ............ To select a digit.
key ............ To select a value.
ENTER key ...... To store the span parameter and
proceed to step 17.
ESC key ........... To proceed to step 17 without
changing the span parameter.
5.2.10. Exiting the Calibration Mode
Step17 Check the CALEnd display.
Use the following keys.
CAL key..... To store the parameters and display
CALoFF . Proceed to the next step.
ESC key .... To store the parameters temporarily.
Proceed to the CAL 0 display.
5.2. Calibration Procedure
Turn on STABLE
Span
Maximum
To step 14, step 16
To step 12
ZERO
TARE
New mass
Mass of
maximum
To step 14
To step 16
Place the displayed
mass on the
weighing unit
step
13
F 2
ENTER
Stored input
New input
mV/V
+/-
ZERO
TARE
F 2
ENTER
mV/V
Turn display off