10.
Code Memory
The
AD-4406A has four code memories (1 through 4). Each code memory stores a
set of setpoints.
The data is stored in non-volatile memory, and is not lost even if power is turned off.
Memory number 0 is a temporary memory and the data is lost when power is
turned off.
The active code memory number can be changed by key switch, external control
input, or a command via the serial interface.
Changing the Code Memory Number by External Input
The code memory number can be changed by external input from OP-05 or
OP-08.
Set F14 and F15 to 18.
F13=18 : Inhibit reading EXT.2 and EXT.3
to prevent unintentional reading when
switching the codes.
EXT2 EXT3 Memory
Number
ON ON
1
OFF
ON 2
ON OFF
3
OFF
OFF 4