AD-4406A
Weighing Indicator
Page
49
Enabling Key Switches
Enables the key switches that are disabled by the DK command. Not applicable to the
keys disabled by F12.
Template
EK,n
n: key switch number (0: all keys, 1 - 6: refer to F-Function F12)
Command
Example: to enable all keys
Reply
13.3.3. Commands to Set Parameters
Set Limit/Setpoint Value
Sets the limit or setpoint value of the comparison. The decimal point is not necessary.
Template
Sm,n, [ value ]
m: code memory number, 0 - 4
n : setpoint order number, refer to "Setting the setpoint"
Command
Reply
Set Zero Band
Sets the F26 value (zero band) of the comparison. The decimal point is not
necessary.
Template
SZ, [ value ]
Command
Reply
13.3.4. Commands for Hold Function
Start Averaging to Hold
Starts averaging to hold. The reply differs with the conditions.
Template
HS
Command
Reply
Averaging
start
Averaging
now
Hold
Release the Hold Data
Releases the hold data or stops averaging and goes to the normal weighing mode.
Template
HC
Command
Reply