30
problem lies with the timer. Replace the timer.
(3) If AC voltage is present at both quick-connections leading from the
timer, measure the resistance of each solenoid coil at the quick-
connect leading to the solenoid valve. It should be about 430-ohms if
120V and 1850-ohms if 220V. If other than this, the coil is either
open or shorted and the solenoid valve should be replaced.
(c)
Check the AC voltage going to and from the RFI filter. If voltage is going
in but not coming out, replace the RFI filter. If both voltages are present,
then the timer is defective. Replace the timer.