23
12.0
PREVENTIVE MAINTENANCE.
12.1
BALLAST BLEED
- All accumulated water must be bled from the ballast tank
at least once every three days during continuous operation. This is done by
operating the manual toggle valve, Figures 1, 2 and 9 (12), located on the
ballast tank. There should be a firm flow of air from the BALLAST BLEED
tubing, usually accompanied by water which can be caught in a suitable
container and discarded. The tank should be drained completely. During
periods of high humidity the ballast tank should be bled every day.
12.2
BLEED SYSTEM
- A weekly routine check should be made for plugging of the
bleed system. When the bleed valve is opened there should be a strong flow of
air from the BALLAST BLEED tubing. If the air flow is weak or no water is
emitted, there is a possibility that there is some blockage within the tubing
leading to the bottom of the tank or the manual valve.
12.3
FILTERS
- All filters should be cleaned and replaced once per month if the
compressor operates continuously. The filters in question are the two inlet
filters on the compressor, Figure 8 (44).
12.4
METHANE REACTOR
- A one or two minute observation of the METHANE
HEAT lamp, Figure 4 (4), should reveal cycling of the lamp. This gives positive
indication of heat control to the methane reactor. A glance at the pyrometer will
also reveal that the unit is at temperature (290
C ± 10
C). If either check
reveals some problem, see Section 13.22 for remedial action.
12.5
COMPRESSOR
(See Section 7.0) - A check of the INPUT PRESSURE
gauge, Figure 4 (3), over a one or two minute period should reveal: (a) the
compressor is operating properly (reaches a maximum pressure of about 80-
psig and cuts off), and (b) the pressure switch/compressor pressure relief
system is operating properly (cycles between 60-psig and 80-psig over a one
minute period). If not, see Section 14.0 for remedial action.
12.6
PURIFICATION REACTOR
- A one or two minute observation of the rotameter,
Figure 4 (10), should show the rotameter ball as remaining constant. If both the
rotameter ball and the OUTPUT PRESSURE gauge, Figure 4 (6), show sudden
drops during a one minute cycle, see Section 13.8 for cause and remedy.
Summary of Contents for 737 Series
Page 4: ...v 20 737 10 ROTAMETER FLOW NOMOGRAPH 62 ...
Page 11: ...7 3 16 Follow the procedures as outlined in Section 3 7 through Section 3 13 ...
Page 36: ...32 defective check valve ...
Page 48: ...44 ...
Page 49: ...45 ...
Page 50: ...46 ...
Page 51: ...47 ...
Page 52: ...48 ...
Page 53: ...49 ...
Page 54: ...50 ...
Page 55: ...51 ...
Page 56: ...52 ...
Page 57: ...53 ...
Page 58: ...54 ...
Page 59: ...55 ...
Page 60: ...56 ...
Page 61: ...57 ...
Page 62: ...58 ...
Page 63: ...59 ...
Page 64: ...60 ...
Page 65: ...61 ...
Page 66: ...62 ...
Page 67: ...63 ...
Page 68: ...64 ...
Page 69: ...65 ...
Page 70: ...66 ...
Page 72: ...68 ...