- 24 - - 24 -24
Rev D
24
Fault
Causes and remedies
Lamp does not start
Check
Top cover is secured. If top
cover is not secured, lamp
will not start.
Check
Lamp life. If 1600 hours
have been exceeded, lamp
must be replaced.
Check
Lamp is intact and properly
installed. If lamp is not
intact or properly installed, it
will not start. Only qualified
personnel can inspect lamp
installation.
Contact
If problem persists, contact
89 North customer service
for further instructions.
Temperature High or Exceeded
Check
Ambient air temp too high,
improve air circulation.
Check
Air vents blocked or
restricted.
Contact
If problem persists, contact
89 North customer service
for further instructions.