4titude Ltd, The North Barn, Surrey Hills Business Park, Damphurst Lane, Wotton, Dorking, Surrey, RH5 6QT, UK
Tel: +44 (0) 1306 884885 Fax: +44 (0) 1306 884886 [email protected] www.4ti.co.uk
Page | 26
The errors will remain displayed until the problem is solved. To clear the
error please retry the operations that generated the error, e.g. error 7
may occur if the drawer is pushed in but released before the motor sets in
motion. To solve the problem please push in the drawer again and hold
until an audible warning is heard.
If any of the above errors persists switch “OFF” the unit and then switch
“ON” again.
If within 10 seconds the problem is not solved the heated plate is
automatically turned off to avoid sample damage.
If the problem is solved within 10 seconds the unit will automatically
return to “Run Mode”.