Page | 19
4.2 Stand-by, Power Down and Audible Warning Control
To programme stand-by and power-down time and control the audible
warning:
•
Press the M key for 10 seconds, then release.
•
The display shows the symbol “L” (flashing) and the last set stand-by
time in minutes. (The stand-by time is the time after which the heater
temperature is automatically reduced to 60°C when the unit is left idle).
•
The stand-by time can be decreased or increased with the
keys
•
Time can be set between 1 minute and 99 minutes with steps of
1minute or as “no” standby.
•
Press the M key again to store the new stand-by time and then the
display will flash with “P” and the current power-down time (in hours).
(The power-down time is the time after which the heater is switched off
automatically when the unit is left idle).
•
The power-down time can be decreased or increased with the
keys
•
Power-down can be set between 1 hour and 24 hours with steps of 1
hour
•
Press the M key again to store the new power-down time and the
display will flash with “S” and the current audible warning setting.
•
The audible warning function can be enabled (“on”) or disabled (“no”)
with the
keys.
•
Press the M key again to store the audible warning function and this
will also exit from the programme menu.
•
The unit returns into “run mode” (see Chapter 5, Basic Functions, page
20) and the actual temperature of the heating platen is shown in the