14
70-2011-5643-0 Issue Date: 10/2015
3M
™
Steri-Vac
™
Sterilizer/Aerator GS Series
Site Planning and Installation Guide
189.2 cm
(74.5 in.)
114.3 cm (45 in.)
188.0 cm (74 in.) Max.
Min. Depth
182.9 cm (72 in.) Min.
111.8 cm (44 in.) Max.
104.1 cm (41 in.) Min.
108.0 cm
(42.5 in.)
114.3 cm
(45 in.)
Figure 7. Typical In-Wall Opening Requirements for 3M
™
Steri-Vac
™
Sterilizer/
Aerator GS Series, Model GS8
Pre-Installation Visit and Site Readiness Verification
Prior to the installation of the 3M
™
Steri-Vac
™
Sterilizer/Aerator GS Series and any other accessory
equipment, the 3M Health Care service personnel or authorized 3M service personnel can be
contacted to visit the facility where the equipment will be installed. The proposed Site Planning and
Equipment Location Worksheets must be reviewed to discuss the required services, equipment
spacing, and any additional considerations. A Site Readiness Verification Form (ANNEX C) must be
completed and documented with the purchaser.