70-2011-5643-0 Issue Date: 10/2015
3M
™
Steri-Vac
™
Sterilizer/Aerator GS Series
Site Planning and Installation Guide
Abator
Indicator
Panel
Abator
Indicator
Panel
EO STERILIZER AND
OPERATOR ROOM
SERVICE
ROOM
GS8
GS5 above
GS5
below
Prep Table
Exhaust Vent
Cart
Flammable
Liquid
Storage
Cabinet
Prep Table
Abator
Abator
10 cm
(4 in.)
CLR
51 cm
(20 in.)
CLR
51 cm
(20 in.)
CLR
51
c
m
(2
0 in.
)
CL
R
51
c
m
(2
0 in.
)
CL
R
51 cm
(20 in.)
CLR
51 cm
(20 in.)
CLR
Pr
ep
Ta
ble
10 cm
(4 in.)
CLR
Proposed Location of Equipment
Using the completed sketch on the Site Planning Worksheet, complete an Equipment Location
Worksheet (ANNEX B) to show the location and size of each piece of equipment in the work area
(Figure 5). Include all other hard good items in the room (e.g. work tables, storage cabinets, etc.),
as this will help to assess the work flow.
After placing the 3M
™
Steri-Vac
™
Sterilizer/Aerator GS Series and any accessory equipment on the
Equipment Location Worksheet, indicate on the Worksheet any additional services required for the new
equipment. Provide the worksheets and service requirements to any contractors (electrical, ventilation,
plumbing, and structural) involved with the project.
Figure 5. Work Area and Equipment Location Example
CAUTION
To reduce patient risks associated with
exposure to potentially non-sterile devices
or inadequate sterilization procedures,
do not
install 3M
™
Steri-Vac
™
Sterilizer/Aerator GS Series
near any device emitting strong electronic magnetic
fields (EMFs).
12