ANNEX D
3M
™
Steri-Gas
™
EO Gas Cartridge Dilution Charts
The following Tables provide estimates of air concentrations of ethylene oxide (EO) if the identified 3M
™
Steri-Gas
™
EO Gas Cartridge should empty into a
room, and provide the time needed for a room ventilation system to reduce the EO concentrations to acceptable levels. These charts are based on rooms
with 10 air exchanges per hour.
3M
™
Steri-Gas
™
EO Gas Cartridge 4-100
3M
™
Steri-Gas
™
EO Gas Cartridge 8-170
Room
volume
(ft
3
)
Room
volume
(m
3
)
Room
airflow
(cfm)
EO gas
weight
(g)
Room EO
concentration
after spill
(ppm)
Time to
reach
0.5 ppm
(minutes)
600
17
100
100
3270
53
700
20
117
100
2803
52
800
23
133
100
2452
51
900
25
150
100
2180
50
1000
28
167
100
1962
50
1500
42
250
100
1308
47
2000
57
333
100
981
45
2500
71
417
100
785
44
3000
85
500
100
654
43
3500
99
583
100
561
42
4000
113
667
100
490
41
4500
127
750
100
436
41
5000
142
833
100
392
40
5500
156
917
100
357
39
6000
170
1000
100
327
39
6500
184
1083
100
302
38
7000
198
1167
100
280
38
7500
212
1250
100
262
38
8000
227
1333
100
245
37
8500
241
1417
100
231
37
Room
volume
(ft
3
)
600
17
100
170
5559
56
700
20
117
170
4765
55
800
23
133
170
4169
54
900
25
150
170
3706
53
1000
28
167
170
3335
53
1500
42
250
170
2224
50
2000
57
333
170
1668
49
2500
71
417
170
1334
47
3000
85
500
170
1112
46
3500
99
583
170
953
45
4000
113
667
170
834
45
4500
127
750
170
741
44
5000
142
833
170
667
43
5500
156
917
170
606
43
6000
170
1000
170
556
42
6500
184
1083
170
513
42
7000
198
1167
170
476
41
7500
212
1250
170
445
41
8000
227
1333
170
417
40
8500
241
1417
170
392
40
Spill ppm = [(gas weight in mg) / (room volume in cu m)] * [(1 ppm / (1.80 mg/cu m)]
cfm = (room volume in cu ft) * (10 air exchanges per hour) / (60 min/hr)
time = [log10(0.5 ppm / spill-ppm)] * [(-2.303) * (room volume in cu m) / (cfm)]
Room
volume
(m
3
)
Room
airflow
(cfm)
EO gas
weight
(g)
Room EO
concentration
after spill
(ppm)
Time to
reach
0.5 ppm
(minutes)
Employee monitoring may be required by national, state and local regulations. In the U.S., the Occupational Safety and Health Administration (OSHA)
requires breathing zone monitoring of employees for exposure to ethylene oxide after the 3M
™
Steri-Vac
™
GS Series Sterilizer/Aerator GS Series installation
to establish a baseline (OSHA 29 CFR 1910.1047). The 3M
™
Ethylene Oxide (EO) Monitoring Badge 3550 can be used to facilitate personal monitoring
compliance to the OSHA standard.
3M
™
Steri-Gas
™
EO Gas Cartridge Safety Data Sheets (SDS) are available at: www.3M.com/msds
50