10
78-8136-3182-3
5. Voltmeter Operation
Voltage Measurement Ranges:
Manual/Auto; 25V, 100V, 500V, 1000V, & 5000V
Meter display symbols in Voltmeter mode
V
= 3M
™
Charge Analyzer 711 in Voltmeter mode (upper left corner).
R
= Voltage measurement range selected. If range selected is too low OVERFLOW will appear.
AR
= Auto voltage measurement range selected (arrow
←
symbol will briefly flash in lower left corner before
symbol appears. Arrow
→
symbol will briefly flash during overflow condition when in the auto range.
U
= Voltage level on cup electrode.