22
78-8136-3182-3
3M
™
Charge Analyzer 711
Verification Check List
Serial Number____________________________________________________________
Date_____________________Inspector___________________________________
Range
Setting
P.S. Voltage
&
Display Voltage
√
(+) Pass (-)
√
(+) Fail (-)
Analog Output
√
(+) Pass (-)
√
(+) Fail (-)
25 VDC
±25Vdc
Full Range
±2.5%
24.4 - 25.6
Full Range
2.0Vdc ±10%
100VDC
±100Vdc
Full Range
±2.5%
97.5 - 102.5
Full Range
2.0Vdc ±10%
500VDC
±500Vdc
Full Range
±2.5%
487.5 – 512.5
Full Range
2.0Vdc ±10%
1,000VDC
±1,00Vdc
Full Range
±2.5%
975 – 1025
Full Range
2.0Vdc ±5%
5,000VDC
±5,000Vdc
Full Range
±2.5%
4875 – 5125
Full Range
2.0Vdc ±10%