D. Surface-to-Groundable Point Test: Test samples per 

Figure 5-4, using both the 100 volts and 10 volts 
SURFACE TEST ranges at 50% RH and record the 
values as (R

TS-GP

). 

Procedure:

Place the Instrument on a table top or other

stable surface. Connect the test leads to the Instrument by
means of the right angle banana plugs. Connect the other
end of one of the leads to one of the test weights and place
the weight on the surface to be tested. Use one of the
supplied clips to connect the other lead to the groundable
point on the static control surface. Depress TEST button for
15 seconds and then record the reading. 

E. Surface-to-surface Test: Test samples as shown in Figure 

5-5; use both test weights and repeat the same test 
procedure used to determine (R

TS-GP

). 

F. Repeat A through E after conditioning samples at 73°F 

(23°C) and 12% RH. Use the same test points and record 
the values.

RESISTIVE CHARACTERIZATION
OF STATIC CONTROL FLOORING

GENERAL

This section provides a summary of installed or applied
floor material measurements specified and described by
ESD-S7.1.  

NOTE: The following paragraphs are offered as a condensed
summary of the test methods and procedures outlined in the
EOS/ESD standard.  For complete details, refer to the standard.

TEST DESCRIPTION

PERIODIC PERFORMANCE TESTING OF INSTALLED
OR APPLIED FLOORING MATERIALS(Measurement of
resistance from the surface of an installed floor to
GROUNDABLE POINT at ambient temperature and
humidity):  Note:  GROUNDABLE POINT is a point on the
floor material that is intended to accommodate an electrical
connection from the floor material to an appropriate
electrical ground.  The ground point may be an electrical
ground, building ground, or other suitable ground.  If  you
have questions concerning the correct ground, refer to
EOS/ESD Standard 6.0 and/or contact a qualified
electrician.

The Resistance-to-Ground test is the most important test
that can be done on a static control surface because it
verifies that the surface is working correctly and will drain a
static charge in a reasonable time.  This test involves
measurement of the total resistance from the static control
surface through the conductor or ground cord to the ESD to
the ESD GROUND (ESDG), verifying that the entire static
control system is functioning correctly.

NOTE:  ESD 7.1 is designed to measure floor materials with
resistances of 2.5 x 10

4

to 1.0 x 10

11

ohms.

The following procedures should be followed when testing
installed static control floor mats or flooring surfaces.

TEST PROCEDURE FOR RESISTANCE 
TO GROUND

A. Complete BATTERY TEST and CONTINUITY TEST.

If required clean, electrodes as described in Test Weight 
Cleaning section page 5.

4

OFF

10V

100V

BATTERY

TEST

CONTINUITY

TEST

TEST

READ MANUAL BEFORE USE

OHMS

SURFACE

TEST

TAUT        BAND

SUSPENSION

10

9

1G

10G

FAIL

PASS

10

10

10

11

100G

10

8

10

7

10

6

10

0

 

1M

100K

0

10M

100M

CONTINUITY

BATTERY

100V

0

10K

100K

1M

10M

100M

OFF

10V

100V

BATTERY

TEST

CONTINUITY

TEST

TEST

READ MANUAL BEFORE USE

OHMS

SURFACE

TEST

TAUT        BAND

SUSPENSION

10

9

1G

10G

FAIL

PASS

10

10

10

11

100G

10

8

10

7

10

6

10

0

 

1M

100K

0

10M

100M

CONTINUITY

BATTERY

100V

0

10K

100K

1M

10M

100M

Figure 5-4: Material Evaluation Surface to Groundable Point Test

Figure 5-5: Material Evaluation Surface to Surface Test

Figure 5-3: Material Evaluation Sample Configuration

2”

2”

TO TEST WEIGHT AT TEST

POINT A & B

TO TEST WEIGHT AT TEST

POINT A & B

STATIC CONTROL SURFACE

GROUNDABLE POINTS

STATIC CONTROL SURFACE GROUNDABLE
POINTS

SAMPLE SIZE: 10” X 24”

10”

10”

STATIC CONTROL SURFACE

GROUNDABLE POINTS

Summary of Contents for 701

Page 1: ...Operator s Manual 701 Test Kit for Static Control Surfaces Megohmmeter...

Page 2: ...either of two test voltages verify proper Instrument performance and check the electrical continuity of the test set up ACCESSORIES All accessories required for operating the Model 701 are included in...

Page 3: ...nds and then read the resistance from the OHMS scale After all readings have been completed return the Main Selector Switch to the OFF position RESISTIVE CHARACTERIZATION OF STATIC CONTROL WORK SURFAC...

Page 4: ...ngs are above 1x109 ohms the static control surface may be dirty Clean the surface using the manufacturer s recommended cleaning procedure followed by two clear water rinses and a rinse with 70 Isopro...

Page 5: ...building ground or other suitable ground If you have questions concerning the correct ground refer to EOS ESD Standard 6 0 and or contact a qualified electrician The Resistance to Ground test is the m...

Page 6: ...f five tests per 5 000 square feet 464 5 m2 of floor material whichever is greater A minimum of three of the five tests should be conducted in those areas that are subject to wear or have chemical or...

Page 7: ...ative humidity up to 90 for altitudes up to 2000m unless otherwise noted Pollution degree 2 class 3 equipment Resistance Ranges Test Measurement Open Circuit Internal Function Range Voltage Resistance...

Page 8: ...rohibited by law 3M will not be liable for any loss or damage arising from this 3M product whether direct indirect special incidental or consequential regardless of the legal theory asserted Shipping...

Reviews: