7. Mechanical Zero Adjust:

This control is used to zero the pointer.

OPERATING INSTRUCTIONS

• Turn off the Instrument before connecting or 

disconnecting test leads, or before moving test weights.

• Do not use this Instrument with any accessories not 

specifically designed to be used with this product.

• Do not use this Instrument to measure live circuits.

• The test weights included in this kit are heavy. Exercise 

care in handling.

NOTE: The following procedures should be followed each time
the Model 701 is used.

BATTERY TEST

Place the Instrument on a table top or other stable surface.
Set the main selector switch to BATTERY TEST. Press
TEST and hold for 15 seconds. The pointer should come to
rest in the green area of the BATTERY scale. If the pointer
is in the red area to the left of 100V, replace the battery and
retest. If the pointer is in the red area to the right of 100V,
the Instrument may need recalibration.

CONTINUITY TEST

Place the Instrument on a table top or other stable surface
and attach the leads as shown in Figure 4-1. Place the test
weights on the calibration plate or other bare metal surface
and plug in the test leads. Set the main selector switch to
CONTINUITY TEST. Press TEST. The pointer should
come to rest in the green section of the CONTINUITY
scale. If not, the test leads may be defective or the weights
may require maintenance or cleaning.

SURFACE TEST (Resistance Measurement)

Refer to the immediate two following sections to determine
which measurement(s) should be used for your application.
Place the Instrument on a table top or other stable surface

and attach the leads as shown in the appropriate sketch
figures 5-1 through 5-5. Set the Main Selector Switch to the
desired SURFACE TEST voltage. Place the test weight(s)
on the surface to be tested and connect the test leads. Press
TEST for 15 seconds and then read the resistance from the
OHMS scale. After all readings have been completed, return
the Main Selector Switch to the OFF position.

RESISTIVE CHARACTERIZATION OF
STATIC CONTROL WORK
SURFACES

GENERAL

This section provides a summary of the types of surface
measurements specified and described by ESD-S4.1.
Measurements are performed for three reasons:

1. Periodic performance testing of installed static control 

work surfaces.

2. Qualification of installed static control work surfaces.

3. Evaluation of static control work surface materials.

NOTE: The following paragraphs are offered as a condensed
summary of the test methods and procedures outlined in the
EOS/ESD standard. For complete details, refer to the standard.

TEST DESCRIPTION

1. PERIODIC PERFORMANCE TESTING

OF INSTALLED STATIC CONTROL SURFACES

(Measurement of resistance from the top of an installed
surface to ESD GROUND (R

TS-ESDG

) at ambient

temperature and humidity): Note: ESD GROUND is the
point at which the ground cord or other grounding
conductor from the static control surface is connected to.
The ground point may be an electrical ground, building
ground, or other suitable ground. If you have questions
concerning the correct ground, refer to EOS/ESD
STANDARD S6.0 and/or contact a qualified electrician.

This Resistance-to-Ground test is the most important test
that can be done on a static control surface because it
verifies that the surface is working correctly and will

OFF

10V

100V

BATTERY

TEST

CONTINUITY

TEST

TEST

READ MANUAL BEFORE USE

OHMS

SURFACE

TEST

TAUT        BAND

SUSPENSION

10

9

1G

10G

FAIL

PAS

S

10

10

10

11

100G

10

8

10

7

10

6

10

0

 

1M

100K

0

10M

100M

CONTINUITY

BATTERY

100V

0

10K

100K

1M

10M

100M

OFF

10V

100V

BATTERY

TEST

CONTINUITY

TEST

TEST

READ MANUAL BEFORE USE

OHMS

SURFACE

TEST

TAUT        BAND

SUSPENSION

10

9

1G

10G

FAIL

PAS

S

10

10

10

11

100G

10

8

10

7

10

6

10

0

 

1M

100K

0

10M

100M

CONTINUITY

BATTERY

100V

0

10K

100K

1M

10M

100M

Figure 4-1: Continuity Test

TEST WEIGHTS

CALIBRATION 

PLATE

2

2

6

1

3

7

5

4

Figure 3-1: Front Panel

Summary of Contents for 701

Page 1: ...Operator s Manual 701 Test Kit for Static Control Surfaces Megohmmeter...

Page 2: ...either of two test voltages verify proper Instrument performance and check the electrical continuity of the test set up ACCESSORIES All accessories required for operating the Model 701 are included in...

Page 3: ...nds and then read the resistance from the OHMS scale After all readings have been completed return the Main Selector Switch to the OFF position RESISTIVE CHARACTERIZATION OF STATIC CONTROL WORK SURFAC...

Page 4: ...ngs are above 1x109 ohms the static control surface may be dirty Clean the surface using the manufacturer s recommended cleaning procedure followed by two clear water rinses and a rinse with 70 Isopro...

Page 5: ...building ground or other suitable ground If you have questions concerning the correct ground refer to EOS ESD Standard 6 0 and or contact a qualified electrician The Resistance to Ground test is the m...

Page 6: ...f five tests per 5 000 square feet 464 5 m2 of floor material whichever is greater A minimum of three of the five tests should be conducted in those areas that are subject to wear or have chemical or...

Page 7: ...ative humidity up to 90 for altitudes up to 2000m unless otherwise noted Pollution degree 2 class 3 equipment Resistance Ranges Test Measurement Open Circuit Internal Function Range Voltage Resistance...

Page 8: ...rohibited by law 3M will not be liable for any loss or damage arising from this 3M product whether direct indirect special incidental or consequential regardless of the legal theory asserted Shipping...

Reviews: