Figure 5-1: Periodic Performance Tests of Installed Surfaces

OFF

10V

100V

BATTERY

TEST

CONTINUITY

TEST

TEST

READ MANUAL BEFORE USE

OHMS

SURFACE

TEST

TAUT        BAND

SUSPENSION

10

9

1G

10G

FAIL

PASS

10

10

10

11

100G

10

8

10

7

10

6

10

0

 

1M

100K

0

10M

100M

CONTINUITY

BATTERY

100V

0

10K

100K

1M

10M

100M

Figure 5-2: Qualification Tests of Installed Surfaces

3

STATIC CONTROL SURFACE

GROUNDABLE POINTS

TO TEST WEIGHT AT TEST

POINT A, B, C, D & E

TO TEST WEIGHT AT TEST

POINT A, B, C, D & E

2”

2”

STATIC CONTROL SURFACE

GROUNDABLE POINTS

drain a static charge in a reasonable time. This test
involves measurement of the total resistance from the
static control surface through the conductor or ground
cord to the ESD GROUND (ESDG), verifying that the
entire static control system is functioning correctly.

NOTE: ESD-S4. 1 suggests that a static control surface that
measures in the range of 1x10

6

ohms to 1x10

10

ohms RTS-

ESDG is acceptable. HOWEVER, 3M RECOMMENDS THAT A
MORE CONSERVATIVE VALUE OF 1X10

9

OHMS BE USED AS

THE HIGHEST ACCEPTABLE RESISTANCE TO ESD GROUND.

The following procedures should be followed when
testing installed static control surfaces. 

A. Complete BATTERY TEST and CONTINUITY TEST.

B. Set the Instrument on a table top or other stable surface 

and place a test weight at the desired test point as shown 
in Figure 5-1. Connect the test leads to the meter using 
the right angle banana plugs at the meter. Then connect 
one of the test leads to the test weight and the other to 
the ESD GROUND using one of the supplied clips.

C. Test the static control surface using the 100 volts 

SURFACE TEST. Depress the TEST button for 15 
seconds, allowing the pointer to stabilize; record the 
readings for each test point. If the reading is below 1x10

6

ohms, check the static control surface for an alternate 
path to ground; correct and retest. If some or all the 
readings are above 1x10

9

ohms, the static control surface 

may be dirty. Clean the surface using the manufacturer’s 
recommended cleaning procedure followed by two 
clear water rinses and a rinse with 70% Isopropyl 
alcohol. Allow the surface to dry overnight and retest. 
If the resistance reading is “infinite,” there is an 
interruption (open) in the ground connection; repair 
and retest.

2. QUALIFICATION OF INSTALLED STATIC

CONTROL SURFACES

[Measurement of resistance of the top surface to the
groundable point of the static control surface (R

TS-GP

)].

GROUNDABLE POINT is the point at which the

grounding conductor is connected to the static control
surface; the GROUNDABLE POINT is most commonly
a snap (mats), a bolt (laminate), or a strip of conductive
foil tape (flooring). This QUALIFICATION
measurement is similar to those described in the test
description section and is used to verify the correct
installation of the GROUNDABLE POINT by the
manufacturer or by the user. While the test procedure is
the same, the test setup is slightly different; see Figure 5-2.

3. EVALUATION OF STATIC CONTROL

MATERIALS

[Measurement of the resistance between two points on
top of a static control surface (R

TS-TS

), and the resistance

between a point on the surface and the groundable point
(R

TS-GP

)]: Material measurements are done to determine

the intrinsic electrical properties of static control work
surface materials. 

NOTE: Values obtained by these tests may not reflect how a
material will perform when installed as a static control surface.

Material evaluations are typically done at two humidities
(12% RH and 50% RH) to determine whether the
electrical properties of the material are humidity-
dependent. If the low humidity test results are borderline
or not within the expected ranges, caution should be
exercised when using such materials in winter months or
in dry conditions. To assure test accuracy, a minimum of
six samples should be tested. The test procedure is as
follows: 

A. Prepare a minimum of six samples of each material 

to be tested as shown in Figure 5-3. 

B. Clean samples to be tested with a 70% Isopropyl alcohol 

solution and condition at 73°F (23°C) and 50% RH for 
48-72 hours. 

NOTE: Samples must be maintained at the appropriate humidity
level throughout the test procedures.

C. Complete BATTERY TEST and CONTINUITY TEST

Summary of Contents for 701

Page 1: ...Operator s Manual 701 Test Kit for Static Control Surfaces Megohmmeter...

Page 2: ...either of two test voltages verify proper Instrument performance and check the electrical continuity of the test set up ACCESSORIES All accessories required for operating the Model 701 are included in...

Page 3: ...nds and then read the resistance from the OHMS scale After all readings have been completed return the Main Selector Switch to the OFF position RESISTIVE CHARACTERIZATION OF STATIC CONTROL WORK SURFAC...

Page 4: ...ngs are above 1x109 ohms the static control surface may be dirty Clean the surface using the manufacturer s recommended cleaning procedure followed by two clear water rinses and a rinse with 70 Isopro...

Page 5: ...building ground or other suitable ground If you have questions concerning the correct ground refer to EOS ESD Standard 6 0 and or contact a qualified electrician The Resistance to Ground test is the m...

Page 6: ...f five tests per 5 000 square feet 464 5 m2 of floor material whichever is greater A minimum of three of the five tests should be conducted in those areas that are subject to wear or have chemical or...

Page 7: ...ative humidity up to 90 for altitudes up to 2000m unless otherwise noted Pollution degree 2 class 3 equipment Resistance Ranges Test Measurement Open Circuit Internal Function Range Voltage Resistance...

Page 8: ...rohibited by law 3M will not be liable for any loss or damage arising from this 3M product whether direct indirect special incidental or consequential regardless of the legal theory asserted Shipping...

Reviews: