background image

0

Fig. 

Fig. 4

Over Range Alarm

An over range alarm is indicated by a flashing display showing the highest value 

in the range of the sensor: 99 ppm for CO (Fig. ). The alarm indicators are the 

same as those for the normal alarm.

Low Battery Alarm

If battery capacity reaches a point where it is no longer sufficient to maintain 

operation of the unit, the LCD will display “bAt” (Fig. 4). Press button to 

acknowledge and silence the alarm. Do not use until battery is replaced by a M 

Service Center.

A calibration verification test (Bump test) should be performed before every use. This is done by checking 

the monitor’s response to a known concentration of certified calibration gas. Calibration of the unit must be 

performed at least every 0 days or immediately if it does not pass a calibration verification test. Sensors used 

beyond the warranty period or exposed to very high concentrations of gas may require more frequent testing.

NOTE:

 The instrument has been calibrated prior to being shipped from the factory. However, to ensure it is 

performing correctly, a calibration or calibration verification test must be performed prior to initial use.

Alarm Mode Table 

Alarm Mode Display

Alarm Sequence

Alarm

Gas Concentration and 

“ALARM”

Repeating Alarm Cycle: 

Three . sec. alarm sequences (vibrate, double beep/double 

flash) followed by five . sec. alarm sequences (double 

beep/double flash). Repeats cycle.

TWA

Gas Concentration 

alternates with “tA”

Repeating Alarm Cycle: 

Three . sec. alarm sequences (vibrate, double beep/double 

flash) followed by five . sec. alarm sequences (double 

beep/double flash). Repeats cycle.

STEL

Gas Concentration 

alternates with “StL”

Repeating Alarm Cycle: 

Three . sec. alarm sequences (vibrate, double beep/double 

flash) followed by five . sec. alarm sequences (double 

beep/double flash). Repeats cycle.

Over Range

“HIGH ALARM” and 

flashing of highest value 

in the range of sensor 

(99 ppm CO)

Repeating Alarm Cycle: 

Three . sec. alarm sequences (vibrate, double beep/double 

flash) followed by five . sec. alarm sequences (double 

beep/double flash). Repeats cycle.

Error

Err

Double beep/double flash every 40 seconds

Fail

FAil

No audible, visual or vibrating alarm

Low Battery

bAt

Double beep/double flash every 40 seconds

Calibration Verification Test and Calibration

 

WARNING

The following steps must be performed when conducting a calibration or calibration verification test 

(bump test) to ensure proper performance of the monitor. Failure to do so may adversely affect product 

performance and

 result in sickness or death.

– Calibrate prior to initial use.
– When performing a calibration or calibration verification test (bump test) only use certified 

calibration gas at the required concentration level. Do not calibrate with expired calibration gas.

– A calibration verification test (bump test) should be performed before each use. 
– If the instrument cannot be calibrated, do not use until the reason can be determined and corrected.
– Do not cover or obstruct display, audible alarm opening or visual alarm cover. 
– Ensure sensor inlet is unobstructed and is free of debris. 
– Ensure calibration hood is removed prior to use.

Performing a Calibration Verification Test (Bump test)

A calibration verification test (bump test) should be conducted every time you use 

the monitor. This is the only way to effectively confirm that all characteristics of 

the monitor and the sensors are working correctly.
Ensure that you are in a non-contaminated environment before performing a 

calibration verification test. To conduct the test, attach the calibration (cal) hood 

on top of the sensor inlet. Ensure the calibration gas matches the sensor installed 

in the instrument. Connect the hose from the gas regulator of the calibration gas 

bottle to the cal hood (Fig. ). Turn on the gas.
Compare the displayed values with those of your reference calibration gas source. 

Apply the calibration gas for a period of at least  -  minutes to ensure sufficient 

response time and steady state readings. If the measurement displayed is within 

±0% of the calibration gas concentration, turn off the calibration gas and 

remove the cal hood. The unit is now ready for use. Otherwise a calibration should 

be performed. 

Fig. 

CO - ppm

110

Performing a Calibration

Ensure that you are in a non-contaminated environment 

before entering calibration mode. From normal mode 

press the button once; the self-test will be performed 

and “CAL” will be displayed (Fig. 6). Press the button 

within  seconds and the unit will enter calibration mode 

indicated by “000” displayed on the LCD (Fig. 7). If the 

unit passed this zeroing step, the display will change to 

“SPn” alternating with “YES” (Fig. 8 & 9). Press button 

to complete calibration with a gas span. Apply gas when 

the display indicates “GAS” alternating with the expected 

gas level (Fig. 0). Apply the appropriate gas ( ppm for 

CO) within 0 seconds. 
Once the gas is sensed the display will read “SPn” 

alternating with the expected gas level. The calibration 

will take approximately  –  minutes. Upon successful 

completion, the display will show the gas reading. Turn off 

the gas and remove the cal hood.
If calibration fails, the “FAIL” message will be displayed 

(Fig. ). Repeat calibration steps. If the instrument still 

fails to calibrate, do not use. 

Do not use the instrument 

until the reason for the “FAIL” message has been 

determined and corrected.

Fig. 6

Fig. 

Fig. 0

Fig. 9

Fig. 8

Fig. 7

Summary of Contents for 110XLS Series

Page 1: ...eur de gaz unique 3M de s rie 110XLS monoxyde de carbone CO Conserver ces directives titre de r f rence Monitor de un Solo Gas Serie 110XLS Instrucciones de uso para el Monitor de un solo gas 3M Serie...

Page 2: ...d result in over exposure and cause sickness or death For proper use see supervisor or User Instructions or call 3M in U S A 1 800 243 4630 In Canada call Technical Service at 1 800 267 4414 Use only...

Page 3: ...tors the ambient air that enters the sensor through the sensor inlet opening by the process of passive diffusion If the level of the target gas detected by the sensor reaches a factory preset alarm po...

Page 4: ...Monoxide CO It is not to be considered all inclusive nor is it intended to replace the policy and procedures for each facility If you have any doubts about the applicability of the equipment to your...

Page 5: ...ckness or death Fig 2 Fig 4 Fig 5 Fig 6 Fig 8 Alarms Viewing Alarm Set Points To display alarm set points press the button 4 times from the normal operating mode The alarm will be displayed Press the...

Page 6: ...g a calibration or calibration verification test bump test only use certified calibration gas at the required concentration level Do not calibrate with expired calibration gas A calibration verificati...

Page 7: ...ation Kit 17L 10 LEL Methane 35 PPM CO 18 Vol OX with Model 713 Fixed Flow 0 2 LPM Regulator and Case 1 Case 933 000 021 3M 4 Gas Mixture Methane Calibration Kit 58L 10 LEL Methane 35 PPM CO 10 PPM HS...

Page 8: ...of a regular maintenance program Any implied warranties arising out of the sale of 3M s Portable Gas Detectors or Monitors including but not limited to the implied warranties of merchantability and fi...

Page 9: ...ATION 22 Mise en marche et arr t de l appareil 23 Lectures de l affichage 23 Mode d utilisation normale 23 Effectuer un auto test 23 Affichage des concentrations maximales 24 Fonction de rappel d talo...

Page 10: ...hage l ouverture de l alarme audible ou le panneau de l alarme visuelle S assurer que l orifice d admission du capteur est d gag et exempt de d bris S assurer que le dispositif d talonnage est retir a...

Page 11: ...able en usine S curit intrins que CSA U classe 1 division I groupes A B C et D T4A Alimentation Pile au lithium de 3 6 volts non rechargeable install e en permanence rempla able en usine Dur e utile d...

Page 12: ...t Pour tout renseignement sur l utilisation ad quate de ce produit consulter son superviseur ou lire les Directives d utilisation ou communiquer au Canada avec le Service technique au 1 800 267 4414 S...

Page 13: ...hage revienne au mode normal La nouvelle valeur de l alarme est maintenant r gl e Alarme de d passement de plage Une alarme de d passement de plage est signal e par le clignotement l affichage de la v...

Page 14: ...probl me n a pas t d termin et corrig Ne pas couvrir ni obstruer l affichage l ouverture de l alarme audible ou le panneau de l alarme visuelle S assurer que l orifice d admission du capteur est d gag...

Page 15: ...intrins que ce qui peut nuire au bon fonctionnement du produit et provoquer des probl mes de sant ou la mort MISE EN GARDE Trousses d talonnage comprenant le gaz indiqu le d tendeur et l tui Pi ce n...

Page 16: ...s OX EX HS et CO 2 ans compter de la date de fabrication Ensembles de capteurs de la s rie 740 2 ans compter de la date de fabrication Tous les autres capteurs 1 an compter de la date de fabrication C...

Page 17: ...e pantalla 39 Modo de operaci n normal 39 Realizaci n de prueba de auto verificaci n 39 Vista de niveles pico 40 Funci n para recordatorio de calibraci n 40 Alarmas 40 Vista de puntos establecidos de...

Page 18: ...ras incorrectas Lo anterior significa que los niveles presentes de gas monitoreado pueden ser mayores y provocar sobre exposici n lo cual pudiera ocasionar alguna enfermedad o incluso la muerte del us...

Page 19: ...te a 450 MHz 5 watts de radio 61 cm 2 pies Teclado Operaci n de un bot n Rango de temperatura de operaci n Ver la tabla a continuaci n Exactitud de sensor despu s de calibraci n Mon xido de Carbono CO...

Page 20: ...e la secci n Calibraci n y verificaci n de calibraci n Despu s de haber completado con xito una prueba de auto verificaci n en la pantalla aparecer el icono check Fig 5 por 24 horas y el equipo cambia...

Page 21: ...antalla regrese al modo normal Se ha activado el nuevo nivel de alarma ADVERTENCIA Abandone de inmediato cualquier ambiente que active la alarma del monitor No hacerlo puede causar enfermedad o muerte...

Page 22: ...ado en el instrumento Conecte la manguera del regulador de gas del contenedor de gas para calibraci n a la campana para calibraci n Fig 15 Abra la v lvula de control de flujo de gas Compare los valore...

Page 23: ...l producto y resultar en enfermedad o muerte 933 000 004 Kit para calibraci n 3M mezcla de 4 gases Pentano 58L 10 LEL Pentano 35 PPM CO 10 PPM HS 18 Vol OX con Regulador modelo 715 de flujo fijo 0 2 L...

Page 24: ...a impl cita de que los equipos son similares a los que se ofrecen en el mercado o garant a de que son id neos para cualesquier otro fin diverso al se alado en este Manual por lo que el usuario es el r...

Page 25: ...ade 55 Leituras do Display Digital 55 Modo de Opera o Normal 55 Realizando um Auto Teste 55 Visualizando N veis de Pico 56 Fun o de Lembrete de Calibra o 56 Alarmes 56 Visualizando os Pontos de Ajuste...

Page 26: ...bump test antes de cada uso Se o instrumento n o puder ser calibrado n o o utilize at que a causa possa ser determinada e corrigida CUIDADO Este instrumento cont m uma bateria de l tio Descarte de ac...

Page 27: ...a Intr nseca CSA EUA Classe I Divis o 1 Grupo A B C D T4A Energia Bateria de l tio de 3 6 volts n o recarreg vel permanentemente montada substitu vel pelo fabricante Vida til da Bateria 2 anos a part...

Page 28: ...sensor Isto deve ser feito atrav s de uma calibra o ou um teste de verifica o da calibra o Consulte a se o Calibra o e Teste de Verifica o de Calibra o destas Instru es de Uso O auto teste ser indicad...

Page 29: ...o 4 vezes a partir do modo de opera o normal Espere 5 segundos at que o display pisque e ent o pressione o bot o repetidamente at que o n vel de alarme desejado seja exibido Espere 5 segundos at que...

Page 30: ...ando o display indicar GAS alternando com o n vel de g s esperado Fig 20 Aplique o g s apropriado 35 ppm para CO dentro de 30 segundos Uma vez que o g s detectado o display indicar SPn alternando com...

Page 31: ...4 Kit de Calibra o 3M Mistura de 3 Gases Metano 34L 10 LEL Metano 35 PPM CO 18 Vol OX com Regulador de Fluxo 0 2 LPM Modelo 713 e Estojo 1 Caixa 933 000 202 Kit de Calibra o 3M Mistura de 4 Gases Pent...

Page 32: ...ecidos acima A 3M n o ser respons vel pela perda do uso de quaisquer de seus produtos ou por custos ou despesas incidentais ou conseq entes ou por danos sofridos pelo comprador ou por qualquer outro u...

Page 33: ...4T1 Division des produits d hygi ne industrielle et de s curit environnementale de 3M Compagnie 3M Canada C P 5757 London Ontario N6A 4T1 3M M xico S A de C V Av Santa Fe No 190 Col Santa Fe Del lvaro...

Reviews: