User Manual
QuantorMed Plus
TM -721-EN
54
①
Contrast: Increases the image’s contrast. The level can be adjusted from 0 to 10.
Higher levels increase the intensity. Filter is not applied when 0 is selected.
②
Detail Contrast: Increases the contrast in regions of the image with low contrast. The
level can be adjusted from 0 to 10. Higher levels increase the intensity. Filter is not
applied when 0 is selected.
③
Edge Enhancement: Enhances edges. It is used to increase visibility of blood vessels
or small calcifications. The level can be adjusted from 0 to 5. Higher levels increase the
intensity. Filter is not applied when 0 is selected.
④
Edge Frequency: Designates the range of the edges to emphasize. The level can be
adjusted from 0 to 5. Lower level emphasizes the smaller edges. Higher levels
emphasize the bigger edges.
⑤
Latitude Reduction: Compresses the low frequency regions to improve the overall
visibility of the image. The level can be adjusted from 0 to 10. Higher levels increase
the intensity. Filter is not applied when 0 is selected.
⑥
Noise Reduction: Reduces the noise in the image. The level can be adjusted from 0 to
10. Higher levels increase the intensity. Filter is not applied when 0 is selected.
⑦
Set: Set the filter values adjusted by the user as the default value for images of the
current region.
⑧
Load: Load and apply filter values of the current region.
⑨
Reset: Reset filters.
⑩
LUT: Go to LUT Curve adjustment screen.
8.10.
Send Image
Immediately transmits current image to the image server designated in Settings. If
[Network/Export
– Options – Show Destination] is checked under Settings, a dialog box for
choosing the destination will be displayed as shown in Figure 8.8 and the user can designate a
destination.