User Manual
QuantorMed Plus
TM -721-EN
23
2.4.4.
Step 3: Scan Blank
Figure 2.18.
Scan Blank
The
Scan Blank
button is enabled after performing the Erase step.
After the Erase step, insert the cassette with no X-ray exposure into the reader and
click
Scan Blank
to acquire the first calibration image.
The mean intensity value of the blank image is displayed beneath the button.
The mean intensity value of the blank image is displayed in green if the value is within
the acceptable range.
The mean intensity value of the blank image is displayed in red if the value is not within
the acceptable range.
If the mean intensity value of the blank image is displayed in red, click
Accept
to
continue anyways or
Reject
to try again.
To lower the mean intensity value of the blank image, click
Erase
again.