10 

 
 

Main Features 

 

State-of-the-Art Scanning Mechanism

 

FireCR Flash

 adopts state-of-the-art scanning mechanisms using high beam delivery efficiency 

optics module to construct its compact and rigid structure. 
 

High Throughput

   

Its unique and patented dual direction scanning mechanism enables to improve the efficiency 
and high throughput. 
 

Scanning Resolution

   

User selectable resolution of 100

μ

m and 200

μ

m allows user to make diagnosis on variable 

purposes. 
 

Detector

 

High sensitive photomultiplier tube equipped in 

FireCR Flash 

delivers high gain, wide dynamic 

range and high speed response for radiographic imaging. 
 

Powerful Acquisition and Diagnostic Software

 

Acquisition and Diagnostic Software designed for the 

FireCR Flash

, and its accurate and rapid 

data processing make the reader more powerful. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Summary of Contents for FireCR Flash

Page 1: ...fications which must be met by 3D Imaging Simulations Corp All information contained herein is intended for guidance purposes only and characteristics of the products and services described in this publication can be changed at any time without notice Products and services may not be available in your local area Please contact your local sales representative for availability information 3D Imaging...

Page 2: ...4 USA Tel 1 703 430 6080 E mail info 3DISCimaging com Contact 815 Tamnip Dong Yuseong Gu Daejeon Korea Tel 82 42 931 2100 Fax 82 42 931 2299 Website www 3DISCimaging com E mail info 3DISCimaging com 3DISC Europe Gydevang 39 41 3450 Alleroed Denmark Tel 45 88 276 650 E mail info 3DISCimaging com ...

Page 3: ... cause harmful interference to radio or television reception which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receiving antenna Increase the separation between the equipment and receiver Connect the equipment into an outlet on a circuit different from that to which...

Page 4: ...tem and accessories the warranty by 3D Imaging Simulations Corp becomes void 3D Imaging Simulations Corp will not accept any responsibility or liability for the proper functioning of the product in such a case DANGER General mandatory action manual The functionality of the system can be destroyed in the case of incorrect use If unauthorized changes have been made to delivered system and accessorie...

Page 5: ...n of this device should only be undertaken by qualified trained personnel The operator should study instructions and precautions carefully before starting to use the device listed here and throughout the manual There are no user serviceable parts inside this device The device should only be opened and serviced by qualified service personnel Failure to heed this warning may result in injury to serv...

Page 6: ...y when the operating instructions are followed Intended Use This device is a Computed Radiography System and intended for use in producing digital X Ray images for general radiography purposes It comprises of reader cassette with reusable imaging plate and workstation software It scans X Ray exposed image plate and produces X Ray image in digital form Then digital image is transferred to workstati...

Page 7: ...t Button 19 3 4 Computer Requirements 19 3 4 1 Recommended Requirement 19 3 4 2 Minimum Requirement 20 3 5 Installation of Acquisition and Diagnostic Software 20 3 6 Connect the Cable and Power Cord 20 3 6 1 Connect Interface Cable 20 3 6 2 Connect the Power Cord 21 3 6 3 Installation Report 21 Chapter 4 Operating 22 4 1 System Specifications 22 4 2 Operation Conditions 23 4 3 Operating Instructio...

Page 8: ...acturer s Declaration Electromagnetic Immunity 28 5 3 Guidance and Manufacturer s Declaration Electromagnetic Immunity 31 5 4 Laser Safety Statement 32 Chapter 6 Warranty and Repair Service 33 6 1 Standard Warranty 33 6 2 Repair Service 33 6 3 Out of Warranty Repair Service 33 6 4 Shipping 34 Chapter 7 Technical Assistance 35 ...

Page 9: ...ny location as a DICOM 3 0 file format in seconds with no loss of resolution Combine the system with reusable phosphor screen equipped FireCR Flash cassettes and portable X ray equipment it can be used for X ray exams everywhere it is needed e g nursing facilities forensic institutions employee screenings etc The FireCR Flash System is ideal for Computed Radiography examinations in busy clinics an...

Page 10: ...m enables to improve the efficiency and high throughput Scanning Resolution User selectable resolution of 100μm and 200μm allows user to make diagnosis on variable purposes Detector High sensitive photomultiplier tube equipped in FireCR Flash delivers high gain wide dynamic range and high speed response for radiographic imaging Powerful Acquisition and Diagnostic Software Acquisition and Diagnosti...

Page 11: ...versal Cassette 25 x 35 containing IP size 24cm x 30cm CR FP 32 001 Universal Cassette 24 x 30 containing IP size 24cm x 30cm CR FP 32 002 Universal Cassette 18 x 24 containing IP size 18cm x 24cm CR FPA 02 001 USB 2 0 Interface Cable CR FPA 02 002 RJ45 CAT 5E FTP Cable 2M Cross type CR FPA 03 00X Power Cord CR FPM 44 001 FireCR Flash User Manual CR FP 35 001 FireKit Optional Product Identificatio...

Page 12: ...y not in USA make sure the power cord is the correct type that is required in your area WARNING Improper disposal of this product may result in environmental contamination When disposing of this equipment contact 3D Imaging Simulations Corp s representative or related organs of government Do not dispose of any part of this equipment without consulting a 3D Imaging Simulations Corp s representative...

Page 13: ...ntaining IP size 24cm x 30cm Universal cassette 24 x 30 containing IP size 24cm x 30cm Universal Cassette 18 x 24 containing IP size 18cm x 24cm USB 2 0 Interface cable RJ45 CAT 5E FTP Cable 2M Cross type FireCR Flash User Manual QuantorMed Plus User Manual QuantorMed Plus Acquisition and Diagnostic Software QuantorMed Dongle ...

Page 14: ...User Manual FireCR Flash TM 401 EN 14 FireKit ...

Page 15: ...kg 47 lbs and requires two persons to lift Carefully place the reader into its final position WARNING Unsuitable Installation Sites Locations with excessive humidity or dust Locations subject to high temperature Locations subject to shaking or vibration Locations exposed to considerable electrical or magnetic noise or other forms of electromagnetic energy Locations with poor heat radiation Free sp...

Page 16: ...ll in a location that is level and stable Installation in an unsuitable location can cause accident or deterioration in image quality WARNING Sliding the reader may result in internal damage or misalignment of the optics External vibration or shock during scanning may affect image quality The scanner must be placed on rigid flat and reinforced desk or tabletop WARNING Do not place anything on top ...

Page 17: ...d or trained personnel should set wall mount installation Free space of 10 cm 4 inch on both sides at least Front Side Back Side Free space of 20 cm 8 inch on back side at least to allow the power switch power cord and USB cord to be reached by hand at all times Free space of 50 cm 20 inch on top side at least for cassette insertion and ejection ...

Page 18: ...mount on a solid wall perpendicular to the floor When attaching to other building materials please contact your local dealer If installed on a slanted wall it may fall and result in severe injury DANGER 3D Imaging Simulations Corp is not liable for product damage or personal injury when the user fails to follow the product installation instruction WARNING Pulling pushing or climbing on the system ...

Page 19: ... and ready to scan Blinking Scanning Blinking Erasing Blinking Blinking Error 3 3 3 Reader Reset Button In case that the reader is stuck during scanning push reader reset button by sharp pin to initialize the reader 3 4 Computer Requirements 3 4 1 Recommended Requirement Operating System Microsoft Windows 7 8 CPU Intel CORE i5 Memory RAM 4 GB or more WARNING Do not push scanner reset button during...

Page 20: ... Acquisition and Diagnostic Software Refer to Acquisition and Diagnostic Software manual 3 6 Connect the Cable and Power Cord 3 6 1 Connect Interface Cable The reader interfaces with computer via USB2 0 cable or Ethernet cable RJ45 CAT 5E FTP 1 Use the USB Ethernet cable in the shipping container 2 Connect the cable to the reader s USB2 0 Ethernet port located on the rear of the reader 3 Connect t...

Page 21: ...onnected to supply mains with protective earth Use only a three wire cord that has grounding This is a safety feature If you are unable to insert the plug into the outlet contact your electrician to replace your obsolete outlet Do not defeat the safety purpose of the grounding type plug WARNING Socket outlet shall be installed near the device and shall be easily accessible Do not place the device ...

Page 22: ...Matrix IP 18cm x 24cm Standard 900 x 1200 High 1800 x 2400 Accepted Cassette Size 35cm x 43cm 25cm x 30cm 24cm x 30cm 18cm x 24cm Gray Scale Resolution 16 bit Eraser Embedded Computer Interface USB 2 0 Ethernet Dimensions 444 x 795 x 134mm Weight 21 5kg Power Requirement 100 240V 50 60Hz System Configuration Tabletop or Wall Mount Image File Format DICOM 3 0 Specifications subject to change withou...

Page 23: ...gree 2 Ingress of Liquids IPX0 Altitude Up to 2 000m Protective Class Class 1 Equipment Maintenance No user maintenance is required and no user service is allowed Please contact technical support if there is a problem Cleaning Do not try to clean inside of the reader Wipe outside of the reader for dust removing with soft and dry cloth DANGER There are no user serviceable parts inside the reader Th...

Page 24: ...the reader Green indication LED is on when the cassette is inserted correctly Insertion direction of the cassette must be parallel to the reader Cassette can be ejected when scanning and erasing are completed Gently pull back the cassette and ejection direction of the cassette must be parallel to the reader 4 3 4 Getting a Scanned Image To getting a scanned image refer to Acquisition and Diagnosti...

Page 25: ...r PSD peak detect Eraser control high voltage generation 9 BLDC Board This parts controls BLDC Motor and square mirror spin 9 RFID Board This part communicates with the RFID tag RFID tag is contain IP Cassette information 9 Peripheral This is peripherals for image acquisition It consists of Eraser which erases residual images in imaging plate Step Motor and Photo Interrupters which moves stage Las...

Page 26: ...e of Manufacture Equipment Power ON Equipment Power OFF Warning Consult Accompany Documents General mandatory action manual General prohibition indication User Manual Reference Directive on Waste Electrical and Electronic Equipment Authorised Representative in the European Community Keep Dry ...

Page 27: ... Flash system should assure that it is used in such an environment Emission test Compliance Electromagnetic environment guidance RF emissions CISPR 11 Group 1 The FireCR Flash system uses RF energy only for its internal function Therefore Its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B The Model FireCR Flash is...

Page 28: ...environment Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial or hospital environment Power frequency 50 60Hz Magnetic field IEC 61000 4 8 3 0 A m 3 0 A m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment...

Page 29: ... P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in meters m Field strengths from fixed RF transmitters as deter mined by an electromagnetic site survey a Should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the foll...

Page 30: ... distance between portable and mobile RF communications equipment transmitters and the FireCR Flash system as recommended below according to the maximum output power of the communications equipment Rated maximum output power W of transmitter Separation distance m according to frequency of transmitter 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 37 0 37 0 74 1 1 ...

Page 31: ... than 3V m a Interference may occur in the vicinity of equipment marked with the following symbol Note 1 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people Note 2 It is essential that the actual shielding effectiveness and filter attenuation of the shielded location be verified to assure that they...

Page 32: ... Radiation above a Class I 1 level during normal Operation user Maintenance or Prescribed Service Condition Wavelength 658 nm Typ Beam Divergence Paraller 9 5 degrees 2 5 2 5 Perpendicular 17 degrees 3 3 Maximum Power of Energy Output 80 mW CW WARNING Never operate or service the product with the protective cover removed from Laser Reader assembly The reflected beam although invisible can damage y...

Page 33: ...alid when the product is used for its intended purpose and does not cover products which have been modified without written permission from 3D Imaging Simulations Corp or which have been damaged by abuse accident or connection to incompatible equipment This warranty is in lieu of all other warranties expressed or implied 6 2 Repair Service The company reserves the right to cease providing repair m...

Page 34: ... an arrow showing which way is up We do not recommend using loose foam pellets to protect the FireCR Flash If the carton is dropped by the shipper there is a good chance that the device will shift within the loose pellet packing and be damaged If user needs to ship the FireCR Flash to another location or back to the factory and user does not have a means to adequately package it user can order add...

Page 35: ...e If user has any questions about installing or using the device contact your 3D Imaging Simulations Corp representative or your local dealer 3D Imaging Simulations Corp 815 Tamnip Dong Yuseong Gu Daejeon Korea Tel 82 42 931 2100 Fax 82 42 931 2299 www 3DISCimaging com ...

Page 36: ...mpleted form signed by customer to Fax 82 42 931 2299 E mail support 3DISCimaging com Date of Installation Customer Information Hospital Institute Name Address Tel Fax E mail Installer Information Company Name Address Tel Fax E mail System Information Model FireCR Flash CR Reader System S N Installer s Signature Date Customer s Signature Date ...

Page 37: ...ibrate the scanner before use The Please calibrate system message pops up if the scanner is not calibrated or calibration data does not exist It is recommended to calibrate the reader after moving it and as a part of regular maintenance WARNING Calibration should be done for both a Universal Cassette containing IP 14 x 17 and a Universal Cassette containing IP 10 x 12 ...

Page 38: ...ion image file Calibration Create a calibration data file After a successful calibration the calibration window will automatically close Cancel Close the IP Calibration window This will abort the current calibration Cassette Size Select the IP size for calibration Accept Accept a calibration image file The Accept button is displayed to accept calibration image files with values out of range Reject...

Page 39: ...User Manual QuantorMed Plus TM 721 EN 3 1 1 1 Calibration Geometry Figure 2 15 Calibration Geometry WARNING X ray radiation field must cover the whole area of the cassette ...

Page 40: ... Figure 2 16 Auto Alignment Perform Auto Alignment before starting calibration to find the optimal laser beam position 1 1 3 Step 2 Erase Figure 2 17 Erase The Erase button is enabled when the cassette is inserted to the reader It erases the cassette before calibration ...

Page 41: ... The mean intensity value of the blank image is displayed beneath the button The mean intensity value of the blank image is displayed in green if the value is within the acceptable range The mean intensity value of the blank image is displayed in red if the value is not within the acceptable range If the mean intensity value of the blank image is displayed in red click Accept to continue anyways o...

Page 42: ...ue of the low dose image is displayed beneath the button The mean intensity value of the low dose is displayed in green if the value is within the acceptable range The mean intensity value of the low dose is displayed in red if the value is not within the acceptable range click Accept to continue anyways or Reject to try again Increase the X ray dose to increase the mean intensity value or lower t...

Page 43: ... acquire the third calibration image The mean intensity value of the mid dose image is displayed beneath the button The mean intensity value of the mid dose is displayed in green if the value is within acceptable range The mean intensity value of the mid dose is displayed in red if the value is not within the acceptable range click Accept to continue anyways or Reject to acquire the mid dose image...

Page 44: ...f the high dose image is displayed beneath the button The intensity value of the high dose is displayed in green if the value is within the acceptable range The intensity value of the high dose is displayed in red if the value is not within the acceptable range click Accept to continue anyways or Reject to acquire the high dose image again Increase the X ray dose to increase the mean intensity val...

Page 45: ...uccessful Scan High Dose Click the Calibration button to generate calibration data Calibration data is saved in your local program folder After generation of calibration data the calibration window closes automatically 1 1 9 Step 8 Cancel You can exit calibration window in the middle of process This will abort your current calibration process ...

Page 46: ...ithin acceptable range values are shown in green Figure 2 23 Proper Calibration When the mean intensity value for each step is not within acceptable range values are shown in red Figure 2 24 Improper Calibration Start the calibration process from Auto Alignment or Erase for recalibration of the reader ...

Page 47: ...User Manual QuantorMed Plus TM 721 EN 11 NOTE Following the calibration process two sets of four images used for calibration and two calibration files are generated in local program folder ...

Page 48: ...e deba cumplir 3D Imaging Simulations Corp Toda la información contenida en este documento se ofrece de forma meramente orientativa y las caracterí sticas de los productos y servicios descritos en esta publicación se pueden modificar en cualquier momento sin previo aviso Es posible que los productos y servicios no estén disponibles en su región Póngase en contacto con su representante de ventas lo...

Page 49: ...re QuantorMed En este manual se utilizarán los siguientes sí mbolos PELIGRO La funcionalidad del software puede quedar dañada si se hace un uso indebido de este Si se llevan a cabo modificaciones no autorizadas en los componentes de software y hardware la garantí a de 3D Imaging Simulations Corp perderá su validez 3D Imaging Simulations Corp no asumirá responsabilidad alguna en cuanto al correcto ...

Page 50: ...llas 7 2 1 Resoluciones compatibles 7 2 2 Inicio 7 2 3 Barra de navegación 8 2 4 Estudios 8 2 5 Nuevo estudio 9 2 6 Selección de zona 9 2 7 Información del paciente 9 2 8 Escaneado 10 2 9 Revisión 10 3 Nuevo estudio 11 3 1 Diseño de la pantalla 11 4 Selección de zona 13 4 1 Diseño de la pantalla 13 5 Imagen 15 5 1 Diseño de la pantalla 15 5 2 ROI 16 5 3 Marcado 17 5 4 Girar Voltear 18 5 5 Ventana ...

Page 51: ...Unir 21 6 Revisión 23 6 1 Diseño de la pantalla 23 6 2 Barra de tí tulo de la vista 25 6 3 Pantalla completa 25 6 4 Manipulación de imágenes 26 6 5 Medición 27 6 6 Marcado 30 6 7 Archivo 30 6 8 Herramientas 31 6 9 Rango de aplicación 31 6 10 Exportación 32 ...

Page 52: ...900 1 1 2 Requisitos informáticos mí nimos Sistema operativo Microsoft Windows XP CPU Procesador Intel Memoria 2 GB de RAM o más Disco duro 80 GB de espacio libre en disco duro Red Ethernet de 1 Gbps Ví deo Pantalla en color de 32 bits Resolución de ví deo 1366 x 768 1280 x 800 1600 x 900 1280 x 1024 1 2 Inicio y cierre 1 2 1 Inicio del sistema 1 Encienda el lector de CR FireCR 2 Encienda el monit...

Page 53: ...clic en el icono de QuantorMed del Escritorio tal como se muestra en la figura 1 1 QuantorMed Figura 1 1 Icono de acceso directo 1 2 3 Cierre del programa Seleccione el botón Sistema de la ventana Inicio tal como se muestra en la figura 1 2 Figura 1 2 Botón Sistema A continuación podrá salir de la aplicación seleccionando la opción Salir del menú Sistema Figura 1 3 Menú Sistema ...

Page 54: ...í nimo mientras que la resolución vertical debe ser de 768 pí xeles como mí nimo En la tabla 2 1 encontrará una lista de las resoluciones recomendadas Tabla 2 1 Relación de aspecto de la pantalla Resolución de la pantalla 16 9 1366 x 768 1280 x 800 1400 x 900 1600 x 900 1920 x 1080 4 3 5 4 1280 x 1024 2 2 Inicio La pantalla de inicio es la primera pantalla que aparece cuando se ejecuta el programa...

Page 55: ...er a la pantalla de estudios haciendo clic en el botón Estudios de la barra de navegación La barra de subnavegación se encuentra en la parte izquierda de la pantalla En la barra de subnavegación puede seleccionar las ventanas Lista de estudios Lista de trabajo o Transporte relacionadas con el estudio 1 Lista de estudios Puede realizar varias tareas buscando los estudios existentes y seleccionando ...

Page 56: ... En la pantalla de selección de zona puede seleccionar la zona que desee escanear y añadirla a la lista de tareas 2 7 Información del paciente La barra de de información del paciente aparece en la parte superior de todas las pantallas a excepción de las pantallas de inicio y de nuevo estudio En la barra de información del paciente se muestra la información del paciente asícomo otros datos relacion...

Page 57: ...Nombre del modelo 8 Sistema 2 8 Escaneado En la pantalla de escaneado puede insertar un casete en el FireCR y adquirir las imágenes 2 9 Revisión En la pantalla de revisión puede consultar las imágenes escaneadas del mismo modo que en el visor de PACS La pantalla de revisión resulta muy útil para realizar diagnósticos después de adquirir las imágenes Figura 2 7 Pantalla de escaneado Figura 2 8 Pant...

Page 58: ...as Los campos obligatorios se rellenan automáticamente Esta función se utiliza en casos de emergencia en los que no se ha identificado la información del paciente para realizar un primer escaneado de este modo los datos del paciente se rellenan más tarde 2 Editar campo de introducción Aparece el cuadro de diálogo Edición de campos de introducción En el cuadro de diálogo Edición de campos de introd...

Page 59: ...er N A hace referencia a otro O Edad Introduzca la edad del paciente Deje el campo Año en blanco si la edad del paciente es inferior a un año Si solo se introduce la edad sin la fecha de nacimiento esta se calculará automáticamente En este caso el dí a y el mes que se introducen automáticamente son el 1 de enero X Médico referidor Introduzca el nombre del médico referidor X Descripción Introduzca ...

Page 60: ...ponda se marcará con un color distinto Seleccione la zona que desee y se actualizará la lista de posiciones de vista 2 Posición de vista Muestra las posiciones de vista que se corresponden con la zona seleccionada en la que realizar el escaneado Haga clic en la zona que vaya a escanear y esta se añadirá a la lista de tareas 3 Predefinido Muestra los ajustes predefinidos de la posición de vista Cua...

Page 61: ...vanzar página Botón para cambiar de página 5 Lista de tareas Lista de posiciones de vista añadidas desde los campos Posición de vista o Predefinido 6 Anterior Le permite volver a la pantalla de nuevo estudio 7 Siguiente Le permite avanzar a la pantalla de escaneado ...

Page 62: ...para recortar la zona válida de la imagen escaneada ROI Herramientas relacionadas con la ROI Marcado Añada marcas en la imagen Girar Voltear Gire o voltee la imagen Ventana automática Ajusta automáticamente el brillo y el contraste de la imagen Repetir Escanea de nuevo la imagen Rechazar Rechaza la imagen CC Muestra herramientas para ajustar los detalles de la imagen Enviar imagen Transmite la ima...

Page 63: ... a la pantalla tal como se muestra en la figura 5 2 Para ajustar el tamaño del cuadro de la ROI arrastre los rectángulos pequeños de cada esquina de la ROI Al transmitir o imprimir la imagen solo se utiliza la ROI Figura 5 2 Imagen con ROI recortada Para que vuelva a aparecer el cuadro de la ROI haga doble clic en la imagen o bien haga clic en el botón ROI Al hacer clic en el botón ROI los siguien...

Page 64: ...Seleccione una marca Haga clic en la imagen Las marcas predefinidas pueden editarse en Configuración debajo de Superponer Marcado Introducción manual de marcas Haga clic en el botón Marcado Seleccione Introducción directa Haga clic en la posición en la que desee añadir la marca y aparecerá un cuadro de introducción de texto Introduzca el texto Haga clic fuera del cuadro de introducción de texto cu...

Page 65: ... necesario volver a escanearla Recuerde que la imagen actual se eliminará 5 7 Rechazar Cuando la calidad de la imagen escaneada es insuficiente la opción Rechazar conserva la imagen actual con una marca de rechazo sin tener que volver a escanearla Cuando se hace clic en el botón Rechazar se añade una marca de rechazo en la esquina superior izquierda de la imagen tal como se muestra a continuación ...

Page 66: ... de inmediato la imagen actual al servidor de imágenes designado en Configuración Si Red Exportación Opciones Mostrar destino está marcada en Configuración aparecerá un cuadro de diálogo para que el usuario puede elegir el destino tal como se muestra en la figura 5 8 Figura 5 8 Selección de destino ...

Page 67: ...rupan por estudios y se ordenan cronológicamente El usuario puede seleccionar la zona que desee y realizar un escaneado La lista de tareas consta de una lista de estudios y cada estudio consta de una lista de tareas Las casillas que se encuentran a la derecha de la barra de tareas se utilizan para seleccionar tareas Figura 5 1 Lista de tareas Barra del estudio Muestra la fecha la hora y la descrip...

Page 68: ... en una cuando se escanean huesos largos Para unir imágenes todas ellas deben tener la misma resolución de escaneado Las imágenes escaneadas en resolución SD y HD no pueden unirse Pueden unirse hasta tres imágenes en una Desplazamiento Cambian el orden de las tareas Cambie el orden de una tarea pulsando los botones arriba y abajo Cambio Cambia las imágenes de dos tareas Este botón solo se activa c...

Page 69: ...o de la pantalla 5 Recortar Muestra u oculta la lí nea de ajuste de recorte de la imagen Las lí neas de recorte aparecen en forma de lí neas discontinuas Arrastre la lí nea discontinua con el ratón para recortar la zona deseada 6 Mostrar lí nea de guí a Muestra u oculta la lí nea discontinua horizontal para comprobar el nivel horizontal de la imagen escaneada Puede ajustarse la posición de la lí n...

Page 70: ...istas Cada vista consta de varias imágenes El usuario puede personalizar los diseños de la vista y de la imagen conforme a sus necesidades Imagen Imagen Vista Imagen Imagen Vista Vista Figura 6 1 Vista Diseño de la imagen Para que se muestre una imagen en la vista simplemente arrastre la imagen desde la barra de tareas Cuando se arrastra una imagen desde la barra de tareas solo se muestra la image...

Page 71: ... hacia la izquierda la derecha arriba o abajo con el ratón Desplazar Desplaza la imagen Arrastre la imagen hacia la izquierda la derecha arriba o abajo con el ratón Manipulación de imágenes Abra las herramientas de manipulación de imágenes para girar voltear e invertir imágenes Medición Ofrece varias herramientas de medición de distancias y ángulos Marcado Ofrece herramientas de marcado Archivo Mu...

Page 72: ...x 3 tal como se muestra en la figura 6 4 aparecerá en la pantalla Seleccione el diseño que desee El diseño de la imagen más grande disponible es de 3 x 3 Figura 6 4 Diseño de la imagen Retroceder página Permite volver a la página anterior Avanzar página Le permite ir a la página siguiente Cerrar Cierra todas las imágenes 6 3 Pantalla completa Expande la pantalla de revisión al modo de pantalla com...

Page 73: ...d Plus TM 721 ES S 26 Figura 6 5 Pantalla completa 6 4 Manipulación de imágenes Icono Acción Girar a la izquierda Girar a la derecha Voltear horizontalmente Voltear verticalmente Restablecer la imagen a su estado inicial Negativo ...

Page 74: ... Distancia Mide la distancia Haga clic en los puntos inicial y final de la medición Aparece una lí nea recta entre los dos puntos y el valor de medición se muestra al final de la lí nea recta Puede desplazar la lí nea de medición arrastrándola Puede desplazar cualquiera de los dos puntos de control arrastrándolo con el ratón También puede cambiar la posición de los valores de medición arrastrándol...

Page 75: ... suelte el botón del ratón Haga clic en la zona interior del rectángulo para desplazarlo Arrastre los puntos de control para cambiar el tamaño del rectángulo Elipse Dibuja una elipse Haga clic con el ratón en la posición en la que desee que el punto superior y el punto del extremo izquierdo de la elipse se crucen arrastre el cursor hacia abajo a la derecha y suelte el botón del ratón cuando haya c...

Page 76: ...lla las coordenadas x y y los valores de pí xeles p de la posición del cursor Mide la proporción entre las áreas cardiaca y torácica 1 Haga clic en el punto final izquierdo del tórax 2 Haga clic en el punto final derecho del tórax 3 Desplace el punto de control superior izquierdo al punto final izquierdo del corazón 4 Desplace el punto de control superior derecho al punto final derecho del corazón...

Page 77: ...nicial Información del punto inicial 3 Punto final Información del punto final 4 Mí n Valor mí nimo 5 Máx Valor máximo 6 Media Valor medio 7 Valor de pí xeles Valor de pí xeles de la posición actual del cursor al hacer clic en el gráfico 6 6 Marcado Icono Acción Flecha Texto 6 7 Archivo Icono Acción ...

Page 78: ...otón Calibración de pí xeles 3 El siguiente cuadro de diálogo aparecerá en la pantalla Introduzca la longitud real de la herramienta de medición 4 Haga clic en el punto inicial de la herramienta de medición sobre la imagen y empiece a arrastrarlo 5 Suelte el botón del ratón en el punto final de la herramienta de medición para completar la calibración 6 9 Rango de aplicación Establece el rango en e...

Page 79: ...ulsada la tecla Ctrl Las imágenes seleccionadas están marcadas con un borde azul Estudio Aplica filtros de imagen a todas las imágenes del estudio que contengan la imagen activa Todas Aplica filtros de imagen a todas las imágenes Los filtros de imagen que pueden utilizarse con Rango de aplicación son los siguientes Zoom Desplazar Girar Voltear Ventana 6 10 Exportación Tabla 6 2 Exportación Disposi...

Page 80: ...M B Mapa de bits Archivo de mapa de bits de Windows C Sin formato Archivo sin formato D Jpg JPEG E Tiff TIFF Mensaje de registro Muestra el progreso cuando se graba en un disco óptico Barra de progreso Muestra el progreso de exportación de las imágenes Lista de estudios Muestra la lista de los estudios seleccionados Lista de imágenes Muestra las imágenes seleccionadas en la lista de estudios Para ...

Page 81: ...r specifications which must be met by 3D Imaging Simulations Corp All information contained herein is intended for guidance purposes only and characteristics of the products and services described in this publication can be changed at any time without notice Products and services may not be available in your local area Please contact your local sales representative for availability information 3D ...

Page 82: ...5 88 276 650 E mail info 3DISCimaging com 3DISC Americas 22560 Glenn Dr Suite 116 Sterling VA 20164 USA Tel 1 703 430 6080 E mail info 3DISCimaging com Contact 815 Tamnip Dong Yuseong Gu Daejeon Korea Tel 82 42 931 2100 Fax 82 42 931 2299 Website www 3DISCimaging com E mail info 3DISCimaging com ...

Page 83: ...ware and hardware components the warranty by 3D Imaging Simulations Corp becomes void 3D Imaging Simulations Corp will not accept any responsibility or liability for the proper functioning oh the product in such a case DANGER General mandatory action manual The functionality of the system can be destroyed in the case of incorrect use If unauthorized changes have been made to delivered system and a...

Page 84: ... password on a pet s name a relative s name or any dictionary word Do not write down the password Do not leave the account logged on User Access Control Configure the workstation to prompt for logon after coming out of stand by mode Internet Usage Accessing to the Internet exposes the computer to a plethora of vulnerabilities such as Viruses Spyware Trojans Hostile Codes It is not recommended to i...

Page 85: ... Manual QuantorMed Plus TM 721 EN 5 There are several vulnerabilities a malicious user could exploit locally Some examples are Theft of equipment Local password cracking Installation of hardware key loggers ...

Page 86: ...rogram 17 2 3 3 Terminate Program 18 2 4 FireCR Calibration 19 2 4 1 Calibration Geometry 21 2 4 2 Step 1 Auto Alignment 22 2 4 3 Step 2 Erase 22 2 4 4 Step 3 Scan Blank 23 2 4 5 Step 4 Scan Low Dose 24 2 4 6 Step 5 Scan Mid Dose 25 2 4 7 Step 6 Scan High Dose 26 2 4 8 Step 7 Calibration 27 2 4 9 Step 8 Cancel 27 2 4 10 Note 28 3 Screens 30 3 1 Supported Resolutions 30 3 2 Home 30 3 3 Navigation B...

Page 87: ... Layout 46 7 2 Exposure Input 47 7 3 Cassette Indicator 47 8 Image 48 8 1 Screen Layout 48 8 2 ROI 49 8 3 Marking 49 8 4 Rotate Flip 50 8 5 Auto Window 51 8 6 Retake 51 8 7 Reject 51 8 8 QA LUT Curve 52 8 9 QA Image Processing Filters 53 8 10 Send Image 54 8 11 Exposure Index 55 9 Task List 55 9 1 Screen Layout 55 9 2 Stitch 57 9 3 Print 58 10 Review 61 10 1 Screen Layout 61 10 2 View Caption Bar ...

Page 88: ...ayout 77 13 2 List 77 14 Settings 79 14 1 Screen Layout 79 14 2 General 79 14 2 1 Option 79 14 2 2 User Information 80 14 2 3 Password for Processing Parameter 80 14 2 4 Display Language 80 14 3 Fonts 80 14 3 1 Common 80 14 3 2 Marking 81 14 3 3 Printer 81 14 4 Network Export 82 14 4 1 Image Server 82 14 4 2 Work List Server 83 14 4 3 DICOM Printer 83 14 4 4 Options 84 14 5 Work List 85 14 5 1 Wor...

Page 89: ...User Manual QuantorMed Plus TM 721 EN 9 14 6 3 Print Overlay Contents 86 14 7 System 86 14 7 1 Exposure Input 86 14 7 2 File Management 86 14 7 3 System Options 87 14 7 4 Date 87 ...

Page 90: ...d the use of the range of facilities included in the software to make the processing and administration of your medical X ray images as efficient as possible Safety Instruction To ensure the safety of patients staff and other persons any changes to software and hardware delivered by 3D Imaging Simulations Corp may only be made with prior written permission from 3D Imaging Simulations Corp Liabilit...

Page 91: ...her adjustments Regions of interest can be highlighted and users can easily change the order of images or delete images in the multi view window Stitching capabilities for images of legs spines and other anatomical areas are particularly useful for orthopedic and chiropractic facilities Integrated Viewer For smaller facilities without a HIS RIS or PACS QuantorMed s local database enables direct ma...

Page 92: ...ard Disk 500 GB Free Hard Disk Space Network 1 Gbps Ethernet Video 32 bit Color Display Video Resolution 1920 x 1080 1440 x 900 2 1 2 Minimum Computer Requirement Operating System Microsoft Windows XP CPU Intel Processor Memory RAM 2 GB or more Hard Disk 80 GB Free Hard Disk Space Network 1 Gbps Ethernet Video 32 bit Color Display Video Resolution 1366 x 768 1280 x 800 1600 x 900 1280 x 1024 ...

Page 93: ...ort Insert Installation CD Install program will automatically launch If the install program is not launched automatically find and execute the QuantorMed 2 x x Setup exe file on the CD Click the Next button as shown in Figure 2 1 Figure 2 1 Welcome Dialog Box Choose all components and click the Next button as shown in Figure 2 2 Figure 2 2 Components Dialog Box ...

Page 94: ...tton as shown in Figure 2 3 Figure 2 3 Components Dialog Box Choose the destination folder and click the Install button as shown in Figure 2 4 Figure 2 4 Installation Location Dialog Box Installation status will be displayed as shown in Figure 2 5 Figure 2 5 Install Status Dialog Box ...

Page 95: ...gure 2 6 Figure 2 6 Installation of dongle Driver Installation status of the FireCR USB driver will be displayed as Shown in Figure 2 7 Figure 2 7 Installation of FireCR USB Driver If installation is successful Figure 2 8 will show up Click Finish to finish the installation Figure 2 8 Installation Finish Dialog Box ...

Page 96: ...User Manual QuantorMed Plus TM 721 EN 16 If the Windows Security Dialog Box pops up after installation click Allow access Figure 2 9 Windows Security Dialog Box ...

Page 97: ...t Program After Windows is successfully booted you can launch the QuantorMed program by double clicking the QuantorMed icon on your desktop as shown in figure 2 11 Figure 2 10 Home Window QuantorMed Figure 2 11 Short Cut Icon WARNING Do not run any application software other than QuantorMed during operation of the scanner This may slow the scanner response ...

Page 98: ...us TM 721 EN 18 2 3 3 Terminate Program Select the System button as shown in Figure 2 12 on the Home Window Figure 2 12 System Button Now you can exit the application by selecting Exit in the System Menu Figure 2 13 System Menu ...

Page 99: ...libration WARNING Calibrate the scanner before use The Please calibrate system message pops up if the scanner is not calibrated or calibration data does not exist It is recommended to calibrate the reader after moving it and as a part of regular maintenance WARNING Calibration should be done for both a Universal Cassette containing IP 14 x 17 and a Universal Cassette containing IP 10 x 12 ...

Page 100: ...tion image file Calibration Create a calibration data file After a successful calibration the calibration window will automatically close Cancel Close the IP Calibration window This will abort the current calibration Cassette Size Select the IP size for calibration Accept Accept a calibration image file The Accept button is displayed to accept calibration image files with values out of range Rejec...

Page 101: ...User Manual QuantorMed Plus TM 721 EN 21 2 4 1 Calibration Geometry Figure 2 15 Calibration Geometry WARNING X ray radiation field must cover the whole area of the cassette ...

Page 102: ... Figure 2 16 Auto Alignment Perform Auto Alignment before starting calibration to find the optimal laser beam position 2 4 3 Step 2 Erase Figure 2 17 Erase The Erase button is enabled when the cassette is inserted to the reader It erases the cassette before calibration ...

Page 103: ...e The mean intensity value of the blank image is displayed beneath the button The mean intensity value of the blank image is displayed in green if the value is within the acceptable range The mean intensity value of the blank image is displayed in red if the value is not within the acceptable range If the mean intensity value of the blank image is displayed in red click Accept to continue anyways ...

Page 104: ...ue of the low dose image is displayed beneath the button The mean intensity value of the low dose is displayed in green if the value is within the acceptable range The mean intensity value of the low dose is displayed in red if the value is not within the acceptable range click Accept to continue anyways or Reject to try again Increase the X ray dose to increase the mean intensity value or lower t...

Page 105: ... acquire the third calibration image The mean intensity value of the mid dose image is displayed beneath the button The mean intensity value of the mid dose is displayed in green if the value is within acceptable range The mean intensity value of the mid dose is displayed in red if the value is not within the acceptable range click Accept to continue anyways or Reject to acquire the mid dose image...

Page 106: ...of the high dose image is displayed beneath the button The intensity value of the high dose is displayed in green if the value is within the acceptable range The intensity value of the high dose is displayed in red if the value is not within the acceptable range click Accept to continue anyways or Reject to acquire the high dose image again Increase the X ray dose to increase the mean intensity va...

Page 107: ...successful Scan High Dose Click the Calibration button to generate calibration data Calibration data is saved in your local program folder After generation of calibration data the calibration window closes automatically 2 4 9 Step 8 Cancel You can exit calibration window in the middle of process This will abort your current calibration process ...

Page 108: ...ithin acceptable range values are shown in green Figure 2 23 Proper Calibration When the mean intensity value for each step is not within acceptable range values are shown in red Figure 2 24 Improper Calibration Start the calibration process from Auto Alignment or Erase for recalibration of the reader ...

Page 109: ...User Manual QuantorMed Plus TM 721 EN 29 NOTE Following the calibration process two sets of four images used for calibration and two calibration files are generated in local program folder ...

Page 110: ...st 1280 pixels and the vertical resolution of the monitor must be at least 768 pixels Recommended resolutions are listed in Table 3 1 Table 3 1 Screen Ratio Screen Resolution 16 9 1366 x 768 1280 x 800 1400 x 900 1600 x 900 1920 x 1080 4 3 5 4 1280 x 1024 3 2 Home The Home screen is the first screen that appears when the program is run From the Home screen you can create a new study or view existi...

Page 111: ... studies You can enter the Studies screen by clicking the Studies button in the Navigation Bar The Sub Navigation Bar is located on the left of the screen In the Sub Navigation Bar you can select a Study List Work List or Transport windows that are related to the study Study List You can perform various tasks by searching existing studies and selecting a study Work List You can query the work list...

Page 112: ... Region Selection In the Region Selection screen you can select the region to scan and add it to the Task List 3 7 Patient Information The Patient Information Bar is displayed at the top of all the screens with the exception of the Home screen and the New Study screen The Patient Information Bar displays patient information and other related information Figure 3 6 Patient Information Figure 3 4 Ne...

Page 113: ... Time Model Name System 3 8 Scan In the Scan screen you can insert a cassette into the FireCR and acquire images 3 9 Review In the Review screen you can query scanned images in the identical manner as the PACS Viewer The Review screen is very useful for conducting diagnosis after acquiring the images Figure 3 7 Scan Screen Figure 3 8 Review Screen ...

Page 114: ...een by using one of the actions below Clicking the Home button in the Navigation Bar Clicking the End Study button in the Scan screen Clicking the End Study button in the Review screen Figure 4 1 Home Window Model Name Displays the model name of FireCR Current Time Current time is displayed in hours and minutes System Displays the System menu Create New Study Go to New Study screen Study List Go t...

Page 115: ...wing menu appears on the screen Figure 4 2 System Menu Settings Displays the Settings window IP Calibration Runs IP Calibration program Scanner Control Runs Scanner Control program Check Update Checks for a new version If a new version is available it is automatically downloaded and installed About this software Displays software version information Exit Exits the program ...

Page 116: ...andatory fields are filled in automatically This function is used in emergencies where patient information has not been identified to perform a scan first and fill in the patient data later Edit Input Field Bring up Input Field Edit dialog In the Input Field Edit dialog you can change the order of the input fields and add or remove input fields System Menu Display the System menu Input Field Detai...

Page 117: ...the Year field blank if the patient is less than 1 year old When only the age is entered and not the date of birth date of birth is calculated automatically When it does month and date are automatically entered as January 1 X Ref Physician Enter the name of the referring physician X Description Enter study description X Accession No Enter accession number Entering an accession number is mandatory ...

Page 118: ...t be removed Figure 5 2 Input Field Edit Input field items to be displayed on the screen Input field items not to be displayed on the screen Move selected items to left and right Items on the left are displayed on the screen and the items on the right are not displayed on the screen Button for switching pages Move the selected items up or down on the list to change order Save changes and close the...

Page 119: ...ion List will be updated View Position Displays view positions that correspond with the selected region for scanning Click on the region to scan and the region will be added to the Task List Add Region View Position Add new region or view position Preset Displays view position presets When a certain setting is used frequently you can easily automatically add multiple view positions by selecting a ...

Page 120: ...kull Lateral Sella Turcica Lateral Facial Bones PA Caldwell Facial Bones Waters Facial Bones Lateral Facial Bones Townes Sinuses PA Caldwell Sinuses Waters Sinuses Lateral Nasal Bones Lateral Mandible AP Mandible Oblique Mandible Lateral Mastoids Laws Mastoids Townes Mastoids Stenvers Temporomandibular Joint Open Mouth Temporomandibular Joint Closed Mouth Thorax Chest PA Chest Lateral Chest Decubi...

Page 121: ...lique Lumbar Spine AP Lumbar Spine Lateral Lumbar Spine Oblique Sacrum AP Sacrum Lateral Coccyx AP Coccyx Lateral Abdomen Abdomen AP Supine Abdomen AP Erect Abdomen Decubitus KUB AP Shoulder Shoulder AP Shoulder Lateral Shoulder Axial Scapula AP Scapula Oblique Scapula Lateral Clavicle AP Upper Extremities Humerus AP Humerus Lateral Elbow AP Elbow Oblique Elbow Lateral Forearm AP Forearm Lateral W...

Page 122: ... Hip Frog Leg Lateral Sacroiliac Joints Oblique Lower Extremities Femur AP Femur Lateral Knee AP Knee Lateral Knee Oblique Knee Skyline Patella Sun Raise Patella Lateral Tibia Fibula AP Tibia Fibula Lateral Ankle AP Ankle Mortise Ankle Oblique Ankle Lateral Foot AP Foot Oblique Foot Lateral Foot Sesamoids Calcaneus Lateral Calcaneus Axial Toes AP Toes Oblique Toes Lateral ...

Page 123: ...u can add delete or edit body part region or position displayed in region selection And Attributes of each preset can be edited Figure 6 2 Add Region View Position Body Part Region Position Add Region Add new region Type the new exam name in the window Figure 6 3 Add Region ...

Page 124: ...he new position name in the window Tick Copy and type existing position in the window to copy properties to the new position Figure 6 4 Add Position Delete Position Edit Position Move Up Move selected position upward Move Down Move selected position downward Close Close window Save Save changes Attribute Open Attribute window for selected position ...

Page 125: ...igure 6 5 Attribute Window Name Name of position ROI Set default ROI Flip Set default flip of the image Rotate Set default rotation of the image Marking Position Set default marking position on the image Marking Set default marking on the image ...

Page 126: ...This function is only used to manually delete the IP Progress The scanning progress is shown in a circle and in percentages Resolution Select the scan resolution HD resolution scans with 100 mm pixel pitch and SD resolution scans at 200 mm pixel pitch Lower pixel pitch results in higher resolution images Cassette Indicator Displays the status of the cassette Exposure Input Enter the X ray exposure...

Page 127: ...A ms mAs 1 O O O Calculated automatically 2 O X X O 3 O O Calculated automatically O 4 O Calculated automatically O O 7 3 Cassette Indicator The Cassette Indicator displays whether the cassette is inserted and recognized correctly Table 7 2 Cassette Indicator Indicator Status A cassette is not inserted A cassette is inserted but cassette size information is not recognized A cassette is inserted ca...

Page 128: ...the valid region of the scanned image ROI ROI related tools Marking Add markings on the image Rotate Flip Rotate flip the image Auto Window Automatically adjust the image brightness and contrast Retake Scan the image again Reject Reject the image QA Bring up tools for adjusting image details Send Image Transmit the image Image Comment Write image comment Exposure Index Show the exposure index valu...

Page 129: ...drag the small rectangles at each corner of the ROI When transmitting or printing the image only the ROI are used Figure 8 2 ROI Cropped Image To bring up the ROI Box again double click on the image or click on the ROI button When you click on the ROI button the film sizes listed below will be displayed on the screen When you select one of the given sizes the ROI Box will be adjusted to that size ...

Page 130: ...Manually Enter Markings Click on the Marking button Select Direct Input Click on the position where you want to add the marking and a text input box will appear Enter your text Click outside the text input box when you are done 8 4 Rotate Flip Used to rotate or flip images Icon Action Rotate Left Rotate Right Flip Horizontally Figure 8 3 Add Predefined Markings Figure 8 4 Manually Enter Markings ...

Page 131: ...ed 8 7 Reject When the quality of the scanned image is insufficient Reject preserves the current image with a Reject marking without rescanning When the Reject button is clicked a Reject marking is added on the upper left corner of the image as shown below and a Reason input field appears on the screen Type in the reason for rejection and click outside the input box when you are done You can cance...

Page 132: ...an be removed by dragging it outside the box Histogram Shows the histogram of the image The bright region at the center is the region currently displayed on the screen The left part of the histogram indicates the frequency where X ray penetration level is low and image is bright The right part of the histogram indicates the frequency where X ray penetration level is high and image is dark Images h...

Page 133: ...rves and window values of the current region Reset Reset curve Processing Go to Image Processing Filters screen 8 9 QA Image Processing Filters Used to adjust image processing filters Adjusting the image requires expert knowledge about images Adjusting the image without proper knowledge may result in lower image quality Figure 8 7 Image Processing Filter Window Center Window Width ...

Page 134: ...he smaller edges Higher levels emphasize the bigger edges Latitude Reduction Compresses the low frequency regions to improve the overall visibility of the image The level can be adjusted from 0 to 10 Higher levels increase the intensity Filter is not applied when 0 is selected Noise Reduction Reduces the noise in the image The level can be adjusted from 0 to 10 Higher levels increase the intensity...

Page 135: ...portional The recommended EI value is between 1800 and 2200 1800 Dose 2200 Dose is optimal Dose 1800 Dose is too low Dose 2200 Dose is too high 9 Task List 9 1 Screen Layout The Task List is a list of scans to be performed Tasks are group by studies and arranged in chronological order The user can select the desired region and perform a scan The Task List consists of a list of studies and each stu...

Page 136: ...yed on the Preset List in the Region Selection screen Delete Delete selected tasks Edit Show hide Move and Switch buttons Send Study Transmit the selected study Destination of the transmission can be added in Settings Print Print the selected images using a DICOM printer or a Windows printer Stitch Used to stitch multiple images into one when long bones are scanned To stitch images all the images ...

Page 137: ...y according to the order in the Task Bar When you adjust the position and angle of the images and begin stitching the two images will be stitched into one and added to the Task List Figure 9 2 Stitch Zoom Used to zoom in and out of the image Drag the mouse over the image to control the zoom It does not zoom in out on a single image and zooms in out on the entire screen Pan Pan the image Drag the m...

Page 138: ...age The position of the guide line can be adjusted by dragging it using the mouse Rotate Left Rotate the selected image counter clockwise Rotate Right Rotate the selected image clockwise Rotate Angle The angle to rotate the image Reset Reset images Cancel Return to the previous screen Stitch Stitch images and return to the previous screen 9 3 Print Windows printers and DICOM printers are supported...

Page 139: ...M film size 8INx10IN 10INx12IN 10INx14IN 8INx10IN 11INx14IN 14INx14IN 14INx17IN 24CMx30CM 24CMx24CM Orientation Select film orientation Portrait Uses portrait orientation Landscape Uses landscape orientation Destination Select DICOM printer Print Start printing Stop Stop printing Cancel Close dialog box When a Window printer is selected the screen changes as shown in Figure 9 4 1 x 1 2 x 2 2 x 1 1...

Page 140: ...User Manual QuantorMed Plus TM 721 EN 60 Figure 9 4 Window Printer ...

Page 141: ... multiple views Each view consists of multiple images The user can customize the view layout and image layout according to their needs Image Image View Image Image View View Figure 10 1 View Image Layout To display an image in the view all you have to do is simply drag the image from the Task Bar When an image is dragged from the Task Bar only the image is displayed in the view When a Study Bar is...

Page 142: ...e Drag the image left right up or down using the mouse Pan Pan the image Drag the image left right up or down using the mouse Image Manipulation Open image manipulation tools for rotating flipping and inverting images Measurement Provides various measurement tools for measuring distance and angle Marking Provides marking tools File Display DICOM file on the screen Tools Provides other tools Apply ...

Page 143: ...e Image Layout button a 3x3 grid shown in Figure 10 4 will appear on the screen Select your desired layout The largest image layout available 3x3 Figure 10 4 Image Layout Page Up Go back to the previous page Page Down Go to the next page Close Close all images 10 3 Full Screen Expands the Review screen to full screen mode as shown in Figure 10 5 ...

Page 144: ...antorMed Plus TM 721 EN 64 Figure 10 5 Full Screen 10 4 Image Manipulation Icon Action Rotate Left Rotate Right Flip Horizontally Flip Vertically Reset image to initial state Negative image Fit to page size Real size ...

Page 145: ...layed between the two points and the measurement value is displayed at the end of the straight line You can move the measuring line by dragging it You can move either one of the two control points by dragging them with your mouse You can also move the position of the measurement values by dragging it with your mouse Angle Measures angle Click on the center point of the angle you wish to measure an...

Page 146: ... and left most point of the ellipse would intersect drag the mouse toward the bottom right and release the mouse after reaching a desired size and shape Click on an area within the ellipse to move it Drag the control points to resize the ellipse Draws a polygon A control point is added every time you click the mouse Double click when you are done drawing the polygon Click on an area within the pol...

Page 147: ...point of the heart 4 Move the upper right control point to the right end point of the heart 5 The measurement value will be displayed in percent Profile Draw a straight line in the same manner as measuring a distance and a profile dialog box will be displayed automatically 1 Distance Distance between two points 2 Start Point Start point information 3 End Point End point information 4 Min Minimum v...

Page 148: ... Icon Action Open DICOM File Open Image File Export as DICOM or image file 10 8 Tools Icon Action Copy to clipboard Copies the active image to the Windows clipboard Sets pixel size 1 Scan a measuring tool with a determined length 2 Select the scanned image and click on the Pixel Calibration button 3 The following dialog box will be displayed on the screen Enter the actual length of the measuring t...

Page 149: ...range in which image processing will be applied When comparing two images you can apply the same filters to both windows for easier comparison Icon Action Active The image that was clicked last is the active image The active image is marked by an orange border Select You can select multiple images by clicking on images while holding down the Ctrl key The selected images are marked by a blue border...

Page 150: ... s study with the selected study s patient Edit Edit the selected studies Delete Delete the selected studies New Study Add a new study Region Open the selected study and go to the Region Selection screen Scan Open the selected study and go to the Scan screen Review Open the selected study and go to the Review screen Send Study Transmit the selected studies Export Export the selected studies to a C...

Page 151: ...g Pages You can switch pages by using the scroll wheel on the mouse or by pressing the Page Up Down buttons Change column width To change column width drag and move the separating line between the column headers The columns in the Study List are described in Table 11 1 Table 11 1 Study List Column Column Title Description Date Study date Time Study time Name Patient name ID Patient ID Sex Patient ...

Page 152: ...to be hidden in the Study List Move selected items left and right Button for switching pages Move the selected items up or down on the list to change order Save changes and close the dialog box Reset the program to default settings 11 4 Search Filter Select the column to search When a filter has not been set and All is selected ...

Page 153: ...the PC will be listed If you select Local you can use other storage devices connected to the PC Label Label to be displayed on disc Media Type Displays the type of the media inserted in the drive Local Folder Activated only when Local is selected in the Device field Select a destination folder to save images Format Select format to save images Supported formats are as follows A DICOM B Bitmap Wind...

Page 154: ... Study List Shows the list of selected studies Image List Shows images selected in the Study List To not export a certain image click and unselect the image Select All Select all images Unselect All Unselect all images Anonymize Delete patient name and ID when saving as DICOM images Close Close the dialog box Abort Stop recording Burn Start recording ...

Page 155: ...ing the Work List is identical to using the Study List Today Only query today s studies Search Query various criteria Using the Work List is identical to using the Study List Waiting Display the list of studies that have not been scanned yet Auto Query Perform an auto query at a preset time Import Import selected items to the Study List Region Open the studies to select and go to the Region Select...

Page 156: ...cription Sch Date Scheduled date Sch Time Scheduled time Mod Modality Name Patient name ID Patient ID Sex Patient sex Birth Patient birth date Age Patient age Station AE Station AE title Station Station Proc Desc Scheduled procedure description Proc ID Scheduled procedure ID Study UID Study instance UID Acc No Accession number ...

Page 157: ...es sent to an image server or a DICOM printer 13 1 Screen Layout Figure 13 1 Transport Today Search only the studies from today Search Search various criteria Search method is identical to Study List Delete Delete selected items Page Up Down Go to another page 13 2 List Column Title Description Date Send date Time Send time ...

Page 158: ...User Manual QuantorMed Plus TM 721 EN 78 Lap Duration in seconds AE Title Destination AE title Host Destination IP address Port Destination port Name Patient name ID Patient ID Status Send status ...

Page 159: ...c item Clicking on the main category item will change the sub category items Clicking on the sub category item will change the specific item in the selected sub category Setting values can be changed in the specific item Main category item list Sub category item list Specific item list 14 2 General 14 2 1 Option Name Description Default Value ...

Page 160: ...ues will be included in the DICOM information 14 2 3 Password for Processing Parameter Name Description Default Value New Password Set password Confirm Password Confirm password Use password to access LUT PROC Activate password for accessing LUT and processing TRUE Use password to save default value Activate password for saving default values in the LUT and processing windows TRUE 14 2 4 Display L...

Page 161: ...me Description Default Value Font Name Name of marking font ARIAL Font Size Size of marking font Adjustable from 16 to 30 points 22 14 3 3 Printer Name Description Default Value Font Name Name of font used to print ARIAL Font Size Size of font used to print Adjustable from 8 to 12 points 10 ...

Page 162: ... Add an image server Name Description Calling AE Title Set AE title of Quantor Called AE Title Set AE title of image server IP Address Image server IP address Port Image server port Description Image server description Edit Edit selected server An identical screen as adding an image server will be displayed Delete Delete the selected server ...

Page 163: ...ork List Server Manages Work List servers Management method is identical to that for image servers Multiple Work List servers can be added but only one can be checked 14 4 3 DICOM Printer Manages DICOM Printers List Management method is identical to that for image servers Unlike adding image servers the following additional settings are required when adding DICOM printers Group Name Description De...

Page 164: ...automatically activated once a server is added Quantor sends images to all checked servers according to the order on the list 14 4 4 Options Name Description Default Value Export To Define the action to take when the Send Study button is clicked You can choose from Send or Print Send Show Destination A Destination list is shown if the Send button is clicked in the Scan screen FALSE Encoding Series...

Page 165: ...ts S Proc ID Keep a column sorted Always show by the sorts when displaying query results to the screen TRUE Use for query conditions Use in query string Some Work List Servers must not use when searching TRUE 14 6 Overlay 14 6 1 Marking This section can be used to edit pre defined markings when adding markings to images in the Scan screen After entering text in the input field you can perform the ...

Page 166: ...ayed on the image Name Description Default Value Exposure Input Do not use kVp mA ms kVp mAs kVp mAs mA kVp mAs ms Do not use 14 7 2 File Management Name Description Default Value Data Path Set the destination folder for images to be stored Installation Folder Database Data Space Allocates disk space for saving images Auto deletion is activated when the allocated disk space is full 30GB Use Auto D...

Page 167: ...et the folder for saving temporary files used in the program Installation Folder Temp Use On Screen Keyboard Decide whether to use the Windows On Screen Keyboard when entering data on the screen FALSE 14 7 4 Date Name Description Default Value Date Format Set the date display format Choose from one of the three formats below YYYY MM DD DD MM YYYY MM DD YYYY YYYY MM DD Separator Set the separator b...

Reviews: