4.
Delete the existing IP address using the backspace button (
1
) on the display of the unit and
enter 5 dots (
2
). The IP address will be restored automatically when the calibration is finished.
5.
Select the check mark on the display of the unit to boot the unit in LCD calibration mode.
6.
Press in the center of the 5 crosshairs to calibrate the display.
7.
The scanner is now ready for use.