Replacing the PMT assembly
1.
Follow step 1 through 9 of
Disassembly
.
2.
Put the scanner onto its side and remove the 6 screws using a 2.5mm hex key and remove the
bottom cover.
3.
Disconnect the BNC (
1
) and HV (
2
) connectors from the back of the main board.
4.
Rotate the unit to access the opposite side and remove the 2 screws fastening the bracket
using a Philips #1 screwdriver. Reuse the bracket during installation.
1
2