42
9.
Remove the 3 screws using a 2.5 mm hex key and remove the motor base.
10.
Remove the drive belt.
11.
Rotate the unit to gain access to the opposite side and unlock (
1
) and disconnect (
2
) the
ribbon cable from the bottom board.
12.
Remove the screws from the bottom board using a 2.5 mm hex key and remove the bottom
board.
1
2