Pannoramic SCAN 150 1.17 User’s Guide
4 Software Description
•
Both for Extended focus and
Z-Stack
scanning methods, you can determine the desired
number of focus levels with the
Focus levels
selection box. The current focus value is the
starting point. If you set
5
for example, one current, two focus levels above and below FOVs
are captured. Therefore, it is advised to choose an odd number for the focus levels. Separate
layers are created and stored this way, and can be opened with Pannoramic Viewer. The Z-
stack slide contains the extended focus layer.
Focus step size
determines the spacing between focus levels. One step is around 0.2 µm.
The default settings are:
Number of focus levels
: 7;
Focus step size
: 5.
4.4.5 Service tab
Figure 43 – Service tab
This tab contains control and maintenance functions.
Emergency Manual Slide Removal Position
The specimen holder will be moved to a position where the slide can be removed manually
after opening the cover (see section
5.3.1 Opening the protective cover
).
Slide Removed Manually
Click this button to notify the software that you have removed the slide manually.
Show images during scanning
If you select this option, you will be able to see the currently scanned image field during
scanning in the live image display field. This slows down digitization process and should be
activated only with purpose.
July 4, 2013 - Rev. 1
3DHISTECH Ltd.
70(130)