Pannoramic 250 1.15 User’s Guide
4 Software Description
If you select the
Interpolated focus
option, the coarse focus level is calculated from three
different focus points of the specimen. Carry out a manual focusing sequence in the
preview image on those three points.
Set the correct focus value of the desired preview position by clicking
Set.
If you want to go
back to any of the previously set focus points, click the appropriate
Goto
button. To delete
the stored values, click
Clear Values
.
The software calculates a coarse focus level from the three values determined this way as a
basis for the fine focusing routine. You can only define the start parameter here. This will,
however, speed up the focusing process.
If you deselect the
Use fine focus
checkbox, the software immediately starts scanning on
the basis of the interpolated values, without pre-fine-focusing.
Manual Exposure
By selecting the
Use manual exposure times
, you can determine the exposure times. The
values are typically set to
Auto
, meaning that the exposure times are determined
automatically. After selecting the corresponding filter and activating the live image, you can
preset the
Exposure value
.
It is recommended to go to various positions in order to determine the intensity distribution
throughout the specimen and define a representative value. You can accept the value for
presetting by clicking
Set Value
. Repeat the procedure for all other filters.
4.5.5 Service tab panel
Figure 62 – Service tab panel
This tab panel contains control and maintenance functions.
Emergency Manual Slide Removal Position
The specimen holder will be moved to a position where the slide can be removed manually
after opening the cover (see section
5.3.1 Opening the protective cover
).
March 13, 2012 – Rev. 1
3DHISTECH Ltd.
94
(
135
)