7
2
Items needed in the field (not included)
•
Screwdriver
•
Wrench(s) (Adjustable)
•
Multimeter
•
Calibration Gases with regulators (flowrate ~ 2.0 lpm)
1.
NO
2
(Nitrogen Dioxide) - 100 ppm
2.
Bar 97 HIGH Blend (Gas Mixture)
NO (Nitrogen Oxide) – 3000 ppm
CO (Carbon Monoxide) – 8%
CO
2
(Carbon Dioxide) – 12%
C
3
H
8
(Hydrocarbon) – 3100 ppm
Routine measurement should be made of
the parSYNC
®
sampling flowrate (with the
standard lengths of hoses when the system
was purchased) and any significant decrease in
flowrate is an indication that the sampling
pump requires replacement (flowrate will
increase with no hose attached).