Note that although their particulate size sensitivities overlap, their optimal particle size and
concentration responsiveness is different between sensors. Thus, particle mass concentration and
number can be extrapolated based on the evaluation of sensor modality and empirical references.
Multi-plex post-processing maps matched up with engine failure response modes may also enable
prediction for engine trouble diagnostic/failed DPF applications.
•
Gaseous Concentration Measurement Sensors (GasMOD™ Cartridge)
Nitric Oxide (NO) - 3-electrode electro-chemical
•
Linear Measurement Range: 0-5,000 ppm NO
•
Filter: Remove effect of SO2 in gas stream
•
Sensitivity: 0.05 ± 0.01 μA/ppm
•
Response Time: <5 Seconds T90
•
Baseline Offset (clean air): 0 to +12 ppm equivalent
•
Zero Shift (0°C to 40°C): <30 ppm
•
Operating Range: -20°C to 50°C
•
Resolution: 1-2 ppm
•
Repeatability: 2% of signal
Nitrogen Dioxide (NO
2
) - 3-electrode electro-chemical
•
Linear Measurement Range: 0-300 ppm NO
2
•
Response Time: <15 Seconds T90
•
Resolution: 0.1 ppm
•
Zero Shift: <0.2 ppm equivalent
•
Operating Range: -20°C to 50°C
•
Repeatability: 2% of signal
•
Linearity: Linear
Non-dispersive Infrared Spectrometer (NDIR) – CO
2
, CO
Solid-state detector
Response Time: <3 Seconds
Operating Range: 0°C to 50°C
NDIR Carbon Dioxide (CO
2
)
Measurement Range: 0-20%
Accuracy: ±0.3% absolute or ±3% relative
Repeatability: ±0.1% absolute or ±2% relative