Pannoramic 1000 2.0 – User’s Guide
6.1.3 Preventive maintenance
6.1.3 Preventive maintenance
To prevent fungus growth
•
Do not leave the equipment in an unconditioned environment with a storage temperature
above 60°C (140°F) or below 0°C (32°F), which may damage the equipment.
•
Maximum operating temperature is 35°C with a relative humidity of less than 75%.
•
Install it in a room with proper air ventilation.
•
Do not let dust or fingerprints remain on optical surfaces.
6.2 Troubleshooting
This section describes the most common troubleshooting solutions for the Pannoramic products. If
any problem arises that does not have solution in this manual, contact the service personnel.
The product does not start
•
Power supply is interrupted
1.
Make sure that the main switch of the internal power supply is on.
2.
Check cable connections.
Virtual slide is improperly focused almost across its entire surface
•
Focus range enabled
•
Disable
Limiting focus range
.
Poor image quality
•
Imaging error (for example, chromatic aberrations)
•
Contact 3DHISTECH Service for support.
Warning!
Do not modify the camera driver settings, as the default factory
settings give the most accurate images.
May 23, 2019 – Rev. 1
3DHISTECH Ltd.
51(79)