Pannoramic 1000 2.0 – User’s Guide
4 Software and Work-flow
◦
Remove Specks
– If selected, the scanner does not digitize objects smaller than the
predetermined size. This helps filtering specks and other uninteresting objects out. Set
the maximum size of specks not to be digitized by adjusting the Speck size slider.
▪
Area Size
– Set the maximum size of specks not to be digitized by adjusting the slider
◦
Dilation
▪
Size –
Adjust the slider to the desired value to enlarge or decrease the scanned area.
The edge is extended discretely by 100 µm each time. The recommended value for
this setting is 1000 µm in Flash scanning mode.
◦
Fill Holes
– If selected, the holes inside the specimen will also be scanned. Often these
holes contain small sample objects that otherwise would not be scanned.
•
Preview Areas
– The preview image shows the area to be digitized with a light coral colored
mask.
◦
Coverslip Area
▪
Coverslip Mode
– Select Auto or Manual
▪
Coverslip Selection
– can be selected in Manual mode only (Width / Height / Area /
Angle)
Figure 20 – Selection box on the Preview image
◦
Selection Area
– If selected, only the area inside the box is scanned. It is useful if you only
want to scan the area that is equivalent with the size of the coverslip, so external specks
will not be scanned.
▪
Grab Compensation Image Inside Area
– On/Off
▪
Selection Area
– Manual settings: Width / Height / Area / Angle
May 23, 2019 - Rev. 1
3DHISTECH Ltd.
41
(