Luna G3 BPAP 25A (G3 B25A) User Manual MID-97/v1.0
36 / 48
Fig. 18-5
(3) Disassemble the PM2.5 Filter from the device, as shown in Fig. 18-6.
PM2.5 Filter
Fig. 18-6
CAUTIONS!
• To avoid material damage, do not place the spare air filter / PM2.5 Filter in direct
sunlight, humid environments, or temperatures below the freezing point. The air filter
/ PM2.5 Filter should be replaced every 6 months (It may be replaced more
frequently based on actual sanitary conditions).
• Operating the device with a dirty air filter may stop it from working properly and
may cause damage to the device.
19. Traveling with the Device
CAUTIONS!
• Empty the water chamber before packing the device for your trip; in order to
prevent any remaining water from entering the device.
•
Using the device at an incorrect elevation setting could result in airflow pressures
higher than the prescribed setting. Always verify the elevation setting when
traveling or relocating.
•
If the device is used when the atmospheric pressure is out of the stated range (See
Section 6), the accuracy of the leakage alert will be affected.
(1) Use the 3B MEDICAL carrying case to carry the device and accessories as a
carry on. Do not put them in your checked baggage.
(2) This device operates on power supplies of 100
~
240 V and 50 / 60 Hz and is
suitable for use in any country in the world. No special adjustment is necessary, but
you will need to find out which power sockets are used in your destination. If
necessary, a power socket adaptor can be purchased in electronics stores.
(3) Remember to bring a spare air filter and the emergency documents (filled and