Luna G3 BPAP 25A (G3 B25A) User Manual MID-97/v1.0
8 / 48
Parameter
Min.
Max.
Unit
Remarks
Frequency
range FDD
Band 4
(1700 MHz)
Uplink
1710
1755
MHz
Module transmit
Downlink
2110
2155
MHz
Module receive
Frequency
range FDD
Band 3
(1800 MHz)
Uplink
1710
1785
MHz
Module transmit
Downlink
1805
1880
MHz
Module receive
Frequency
range FDD
Band 2
(1900 MHz)
Uplink
1850
1910
MHz
Module transmit
Downlink
1930
1990
MHz
Module receive
WARNING!
•
All other wireless technology emitters must be kept at least 2 meters from the
Cellular Module.
7. Available Therapies
The device delivers the following therapies:
CPAP
–
Delivers Continuous Positive Airway Pressure; CPAP maintains a constant
level of pressure throughout the breathing cycle.
AutoCPAP
–
Delivers CPAP therapy and provides an air pressure no less than the
prescribed pressure
based on the patient’s n
eeds.
S
–
A bi-level mode which responds to both inhalation and exhalation by increasing
pressure when you start to inhale and decreasing pressure when you start to exhale.
There is no automatic delivery of a breath you do not inhale. IPAP (Inspiratory
Positive Airway Pressure) and EPAP (Expiratory Positive Airway Pressure) are preset by
home care provider.
AutoS
–
A bi-level mode which responds to both your inhalation and exhalation. The
differential pressure of IPAP and EPAP are presetted by home care provider. While
working in auto feature, the device will automatically adjust the IPAP and EPAP if it
detects an apnea.