12. First Time Setup
12.1 Placing the Device
Place the device on a firm, flat surface.
WARNINGS!
•
If the device has been dropped or mishandled, if the enclosure is broken, or if
water has entered the enclosure, disconnect the power cord and discontinue use.
Contact your home care provider immediately.
•
If the room temperature is warmer than 95°F (35°C), the airflow produced by the
device may exceed 109.4°F (43°C). The room temperature must be kept below 95°F
(35°C) while the patient uses the device.
CAUTIONS!
• Always ensure that the device is placed in an area where the screen and
indicators are clearly visible.
• If the
device has been exposed to either very hot or very cold temperatures, allow
it to adjust to room temperature (approximately 2 hours) before beginning setup.
• Make sure the device is away from any heating or cooling equipment (e.g.,
forced air vents, radiators, air conditioners).
• The device is not su
itable for use in high humidity environments. Make sure that no
water enters the device.
• Make sure that bedding, curtains, or other items are not blocking the filter or vents
of the device.
• Keep pets, pes
ts or children away from the device to avoid small objects being
inhaled or swallowed.
• To
avoid explosion, this device must not be used in the presence of flammable
gases (e.g. anesthetics).
• Tobacco smoke may cause tar build
-up within the device, leading to the malfunctioning of
the device.
• Ai
r must flow freely around the device for it to work properly.
12.2 Installing the Air Filter and Filter Cap / PM2.5 Filter
(1) Attach the air filter to the filter cap, as shown in Fig. 12-1.
Filter Cap
Air Filter
Fig. 12-1