Atlas is 1.70 m tall, weighs 13 kg and comes in two different skin tones. It has a battery life of 8 to 10 hours. The charge status is color-coded and shown 

on the on/off button. Battery status indicator on the power button: white light (100%-50%) green light (49%-15%) red light (14%-5%) red blinking (less than 

5%) . Atlas is charged via the adapter provided and can be used wirelessly during training sessions. 

•  Wear gloves as required during simulation scenarios.

•  Avoid using colored plastic gloves, as they may cause 

discoloration. 

 

•  To maintain Atlas skin, wash hands before use and place 

Atlas on a clean surface.

•  To avoid permanent staining avoid the use of ink pens,  

felt-tipped markers, acetone, or iodine.

•  Avoid placing Atlas on newsprint or colored paper. 

•  After using Atlas, remove glue residue from Atlas skin with 

70% ethyl alcohol wipes. 

•  Ensure that Atlas is properly secured during transportation 

to prevent personal injury or damage to the product.

•  Always lubricate (silicon oil) the airway device (tracheal 

tubes, iGel) before inserting it in the airway. 

 

•  The device may only be stored and charged in dry rooms.

•  Operating temperature: -10 °C up to +35 °C

•  Storage temperature: -10 °C up to +35 °C

•  Air humidity: 15% - 80% 

•  Only use the charger provided by the manufacturer. 

•  Charger connection specifications:  Input 100-240 V

 

Output 50-60Hz

 

12.6V = 1000mA

•  Recommendation: Once the battery is fully charged (the 

light on the charger turns green), take it out of the charger. 

Getting the manikin overcharged puts more stress on the 

battery and can shorten the battery life 

•  Follow the instructions and keep in mind the warnings while 

using the product.

Before using Atlas keep in mind these safety instruction 

i

SAFETY INSTRUCTIONS

To ensure optimum performance and durability of Atlas, read and follow the user information and warnings:

DO NOT

 perform mouth-to-mouth or mouth-to-nose rescue breathing on Atlas.

DO NOT

 introduce fluids into Atlas with the exception of IV/IO systems built into the arms.

DO NOT 

use electrotherapy or a live defibrillator or AED.

DO NOT

 use sharp objects in the airways.

DO NOT

 create a surgical airway.

DO NOT

 perform a relief puncture or inserting thoracic drains.

DO NOT

 use if there are signs of electrical malfunction, or unusual smell of smoke.

DO NOT

 use oxygen. 

DO NOT

 use damaged chargers. 

DO NOT

 take the battery apart. Never try to take the battery out of the device.

DO NOT

 try to repair the device yourself.

DO NOT

 use live defibrillators.

DO NOT

 use live cardiac pacing.

Ø 

Ø

Ø

Ø

Ø 

Ø 

Ø 

Ø 

Ø

Ø

Ø

Ø

Ø

Summary of Contents for ATLAS THE ALS SIMULATOR P75

Page 1: ...ATLAS THE ALS SIMULATOR P75 USER MANUAL 1023796 1024113...

Page 2: ...Safety Instruction 5 Installation 5 Airway 6 Cardiopulmonary resuscitation CPR 8 Circulation 9 Network Router Settings 11 How to connect Atlas with REALITi360 11 How to change the Wi Fi setting on At...

Page 3: ...s for advanced life support in simulation based training and further training DELIVERY CONTENT 1 2 3 4 5 6 7 8 9 11 10 13 1 Consumable bag 2 Atlas charger 3 Tape 4 Lubricant 5 Adapter 6 IV cannulation...

Page 4: ...the monitor interface selected CPR dashboard screen for viewing all relevant parameters and values of cardiopulmonary resuscitation according to the AHA ERC Guidelines Atlas allows observation and rec...

Page 5: ...stored and charged in dry rooms Operating temperature 10 C up to 35 C Storage temperature 10 C up to 35 C Air humidity 15 80 Only use the charger provided by the manufacturer Charger connection specif...

Page 6: ...mulated dry aspiration removing inserted foreign bodies and inserting Guedel tubes OPA or Wendl tubes NPA When single assistant or dual assistant bag mask ventilation BVM is performed correctly on Atl...

Page 7: ...eal intubation Supra glottic airway alternatives like iGel Laryngeal Masks LMA and Laryngeal Tubes LT can be used in accordance with manufacturers instructions Moreover a gastric probe can be inserted...

Page 8: ...rrying out manual thoracic compressions and continuing resuscitation measures The CPR dashboard screen and Generic Defibrillator gives you feedback on the hand position With the use with of REALITi360...

Page 9: ...top of the Atlas s left arm We recommend the use of drilling systems for this procedure When performing administration of medication via the IV pad and or IO access site use a drain bag which should...

Page 10: ...and place the new training pad in the slot At the end of the training session we recommend drying removing the cannulas and taking the drainage system apart When the day s sessions are done flush the...

Page 11: ...Control iPad to the Wi Fi network Both iPads must be on the same Wi Fi network iSimulate 5 0 If the mini router PIX Link is not included in the delivery content set up the Wi Fi net work with the REA...

Page 12: ...hat appears select the corresponding REALITi monitor to your REALITi ecosystem If the monitor does not appear in the list check that the Control iPad and the Monitor iPad Pro are both con nected to th...

Page 13: ...om the main menu on the Monitor iPad 10 On the Monitor iPad Pro the monitor selected from step 9 should appear Start the defibrillator 11 Swipe in from the right screen edge of the monitor iPad show t...

Page 14: ...14 ATLAS THE ALS SIMULATOR P75 Connect Atlas with REALITi360 14 Compress the manikin You should be able to see the CPR compressions in the ECG waveform and the bar graph 13 The connected manikin will...

Page 15: ...the delivery content of Atlas 2 Connect the REALITi Monitor iPad Pro to the Wi Fi network iSimulate 5 0 If the mini router PIX Link is not included in the delivery content set up the Wi Fi network wi...

Page 16: ...e router of your choice Press the arrow button to save Please wait for 30 seconds Turn off the manikin and then power on again Now connect both iPads to the Wi Fi network of your choice The mini route...

Page 17: ...t Multiple network connections may interfere with transmissions between REALITi and Atlas Forget other networks on both your iPads beside the network required to connect with Atlas Default network is...

Page 18: ...meral IO replacement light skin dark skin 1024050 1024347 IV cannulation pad forearm light skin dark skin 1024047 1024345 IV cannulation pad hand light skin dark skin 1024048 1024344 Lung bag 1024056...

Reviews: