ATLAS THE ALS SIMULATOR P75
5 ATLAS THE ALS SIMULATOR P75
| General Handling | Safety Instructions | Installation
GENERAL HANDLING
INSTALLATION
Atlas is 1.70 m tall, weighs 13 kg and comes in two different skin tones. It has a battery life of 8 to 10 hours. The charge status is color-coded and shown
on the on/off button. Battery status indicator on the power button: white light (100%-50%) green light (49%-15%) red light (14%-5%) red blinking (less than
5%) . Atlas is charged via the adapter provided and can be used wirelessly during training sessions.
• Wear gloves as required during simulation scenarios.
• Avoid using colored plastic gloves, as they may cause
discoloration.
• To maintain Atlas skin, wash hands before use and place
Atlas on a clean surface.
• To avoid permanent staining avoid the use of ink pens,
felt-tipped markers, acetone, or iodine.
• Avoid placing Atlas on newsprint or colored paper.
• After using Atlas, remove glue residue from Atlas skin with
70% ethyl alcohol wipes.
• Ensure that Atlas is properly secured during transportation
to prevent personal injury or damage to the product.
• Always lubricate (silicon oil) the airway device (tracheal
tubes, iGel) before inserting it in the airway.
• The device may only be stored and charged in dry rooms.
• Operating temperature: -10 °C up to +35 °C
• Storage temperature: -10 °C up to +35 °C
• Air humidity: 15% - 80%
• Only use the charger provided by the manufacturer.
• Charger connection specifications: Input 100-240 V
Output 50-60Hz
12.6V = 1000mA
• Recommendation: Once the battery is fully charged (the
light on the charger turns green), take it out of the charger.
Getting the manikin overcharged puts more stress on the
battery and can shorten the battery life
• Follow the instructions and keep in mind the warnings while
using the product.
Before using Atlas keep in mind these safety instruction
i
SAFETY INSTRUCTIONS
To ensure optimum performance and durability of Atlas, read and follow the user information and warnings:
DO NOT
perform mouth-to-mouth or mouth-to-nose rescue breathing on Atlas.
DO NOT
introduce fluids into Atlas with the exception of IV/IO systems built into the arms.
DO NOT
use electrotherapy or a live defibrillator or AED.
DO NOT
use sharp objects in the airways.
DO NOT
create a surgical airway.
DO NOT
perform a relief puncture or inserting thoracic drains.
DO NOT
use if there are signs of electrical malfunction, or unusual smell of smoke.
DO NOT
use oxygen.
DO NOT
use damaged chargers.
DO NOT
take the battery apart. Never try to take the battery out of the device.
DO NOT
try to repair the device yourself.
DO NOT
use live defibrillators.
DO NOT
use live cardiac pacing.