25
500°C
500°C
DC 9.2
2.5.5 Last error message
The last error message displayed by the
oven can be called up again here. This is
retained until the oven displays an error
message again and is then updated in
the process (image 25.1).
2.5.6 Muffle operating hours
The operating hours in the individual tem-
perature ranges can be viewed here. The
information relates to the amount of time
time in the respective temperature range
in hours (image 25.2).
2.5.7 Start cleaning program
If you suspect the heating chamber is
contaminated, run the cleaning program.
Impurities are burned out at a very
high temperature (1150°C). Never use
additives such as activated carbon
(graphite pellets) to clean the heating
chamber!
2.5.8 Chamber drying program
(duration: approx. 30 minutes.)
This program is designed to dry the heating
chamber insulation.
It is recommended to let this program run
after a long vacation period or if the device
has not been used for a longer period of
time. Here, the furnace holds at 600°C. The
heating chamber is not completely closed
during this program. The program is part of
the initial installation process.
Display, Status: Settings / Last error message
V-01
Abort: Error vacuum
DC 9.2 Dentin - 10.05.2022 - 14:12
Settings
Muffle hours
Display, Status: Settings / Muffle hours
Enter
Error code: Run_Err: 0x00000100
720°C
Total hours
32
300°C - 399°C
1
900°C - 999°C
1
400°C - 499°C
2
1000°C - 1099°C
0
500°C - 599°C
1
Über 1100°C
0
600°C - 699°C
1
700°C - 799°C
1
800°C - 899°C
1
under 299°C
25
Image 25.1
Image 25.2