17
G
UIDANCE AND
M
ANUFACTURER
’
S
D
ECLARATION
Warning!
Equipment is not suitable for use in the presence of flammable mixtures.
Medical Electrical Equipment needs special precautions regarding Electromagnetic Compatibility (EMC) and
needs to be installed and put into service according to the EMC information provided.
Portable and Mobile RF Communications Equipment can affect Medical Electrical Equipment.
The use of accessories, transducers, and/or cables other than those specified, with the exception of those
sold by the manufacturer as replacement parts for internal components, may result in increased emissions
or decreased immunity of the equipment or system.
The equipment or system should not be used adjacent to or stacked with other equipment. If adjacent or
stacked use is necessary, the equipment or system should be observed to verify normal operation in the
configuration in which it will be used.
M
EDICAL
D
ISCLAIMERS
Warning!
You should not start this or any other exercise program without first consulting your
doctor. To protect yourself from injury, you must read all of the warning information contained in the
Zona Therapy User Guide before using the Zona Plus™.
S
PECIFICATIONS
Power Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.2 Volt LiFePO
4
rechargeable battery (not user replaceable)
AC Adaptor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . (Phihong) EU: PSM03E-050Q; UK: PSM03K-050Q; USA: PSM03A-050Q
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100 – 240V ~ 100mA; 50-60 Hz 6-8 VA
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5V
200mA-1A (USB-A RECP.)
Storage & Transportation Temperature Range. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . -10° to 50°C (14° to 122°F)
Storage & Transportation Humidity Range. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5% to 50% RH non-condensing
Classification with respect to protection from electrical shock. . . . . . . . . . . . Class II (when connected to the charger)
Mode of Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Continuous
Operator Profile: . . . . . . . . . . . . . . . . . . . . . . . . . . . . Any adult including those with a maximum of 8 years of education
IP Code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IP22
Atmospheric Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 800hPa -1060hPa
S
YMBOLS