9650-0536-01 Rev. B
ZOLL
SurePower Battery Pack
Guide
7
Understanding Battery Faults
The
SurePower Battery Pack
can detect one or more simultaneously occurring faults that
render the battery unfit for use.
To determine the faults that affect a battery, follow this procedure when the fault light is on:
The following table lists battery temperature faults and actions you can take to clear these fault
conditions. All other fault codes require you to remove the battery from service. For more
information on these fault codes, see the
SurePower Charger Station
Service Manual.
Press and hold down the
Display
button for
at least
15 seconds.
The
Fault
LED will blink to indicate the code that identifies
the battery fault (there may be a brief delay before the LED
blinks).
The fault codes number from 1 to 29. To indicate fault code
1, the
Fault
LED blinks once. To indicate fault code 2, the
LED blinks twice, and so on.
If the battery detects multiple battery faults, it displays the
first fault code, pauses briefly, and then displays the next
fault code, and so on.
Fault ID
Fault Type
Recommended Action
13
Battery Over Temperature
During Charge.
The battery is at or above its
maximum temperature for
charging (140°F/60°C).
Move the battery to a cooler room or out of
direct sunlight, if possible, and allow the
battery to cool before continuing to charge the
battery.
If the ambient temperature was low when the
battery reached or exceeded its maximum
temperature, do not use this battery.
14
Battery Under Temperature
during Charge.
The battery is at or below the
minimum temperature for
charging (32°F/0°C).
Move the battery to a warmer room (above
32°F / 0°C) and allow the battery to warm up
before charging again.