Displaying Battery Runtime

The 

SurePower Battery Pack

 can display its present status.

Note:

One or more of the LEDs will light automatically when current is flowing into or out 

of the battery.

To display the status of the 

SurePower Battery Pack

, press and 

release the 

Display

 button (momentary press) on the top of the 

battery.

When you press and release the 

Display 

button, all LEDs blink to 

indicate that they work. The battery then lights specific LEDs for 
approximately 10 seconds to indicate the battery’s run time, calibration 
status, and fault state.

We describe what these LEDs indicate in the following sections.

SurePower Battery Pack

 Icon

LED Indications

This is the 

runtime indicator 

that displays the estimated 

runtime of the 

SurePower Battery Pack

 in your defibrillator. 

When you momentarily press and release the 

Display

 

button on the 

SurePower Battery Pack

, the LEDs above 

these icons light for approximately 10 seconds to indicate 
the battery’s estimated run time. 

If the battery is fully charged, the LED over each of the five 
runtime indicator icons illuminates. As the battery 
discharges, the number of illuminated LEDs decreases.

As the 

SurePower Battery Pack

 discharges, pressing the 

Display

 button will light, from left to right, the LED’s that 

indicate the battery’s run time. For example, if four LED’s 
light, the battery can power a defibrillator for approximately 
1 hour and 30 minutes (displayed as 

1:30+

); if three LED’s 

light, the battery can power the defibrillator for approximately 
1 hour (displayed as 

1:00+

), and so on.

When a battery is installed in a ZOLL defibrillator, the 
runtime indicator displays the run time for the battery in that 
specific defibrillator.

When a battery is 

not

 installed in a ZOLL defibrillator, the 

runtime indicator displays the run time for the battery, at a 
nominal drain rate of 2 Amperes (equivalent to a defibrillator 
with most options installed).

Note:

If the left-most LED of the runtime indicator 

blinks, the battery has less than fifteen minutes 

of use left, and defibrillator shutdown will soon 

occur. Recharge the battery as soon as possible.

Содержание SurePower

Страница 1: ...SurePower Battery Pack Guide 9650 0536 01 Rev B...

Страница 2: ...t ZOLL Medical Corporation to determine if additional product information updates are available ZOLL is a registered trademark of ZOLL Medical Corporation SurePower is a trademark of ZOLL Medical Corp...

Страница 3: ...fety hazard Should the SurePower Battery Pack be dropped inspect and test the unit prior to further use following the procedures described in this manual Avoid prolonged exposure of the SurePower Batt...

Страница 4: ...enance instructions If a problem occurs obtain service immediately Do not use alcohol or ketones MEK acetone etc to clean ZOLL rechargable defibrillator batteries Do not sterilize or autoclave ZOLL ba...

Страница 5: ...lows it to provide the following information The battery s run time or capacity The battery s calibration status Battery Faults Indications Figure 1 1 ZOLL SurePower Battery Pack Note ZOLL ships the S...

Страница 6: ...proximately 10 seconds to indicate the battery s estimated run time If the battery is fully charged the LED over each of the five runtime indicator icons illuminates As the battery discharges the numb...

Страница 7: ...ack Icon LED Indications This is the Recalibration icon This LED lights for approximately 10 seconds after you press and release the Display button if the battery needs to be calibrated If the Recalib...

Страница 8: ...onds While you hold down the Display button the runtime indicator displays the battery runtime When you release the Display button the runtime indicator displays the battery capacity for 2 seconds the...

Страница 9: ...ore the LED blinks The fault codes number from 1 to 29 To indicate fault code 1 the Fault LED blinks once To indicate fault code 2 the LED blinks twice and so on If the battery detects multiple batter...

Страница 10: ...6 5 Length x 5 7 Width x 4 4 Height cm Weight 1 7 lb 77 kg Voltage 10 8 V Capacity 5 8 Ah Operating Temperature 32 F to 122 F 0 C to 50 C Optimal Charging Temperature 59 F to 95 F 15 C to 35 C Short T...