background image

9650-0536-01 Rev. B

ZOLL

 SurePower Battery Pack

 Guide

1

Introduction 

The 

SurePower

 

 Battery Pack 

is a rechargable, sealed lithium ion battery for use in 

compatible ZOLL defibrillators and chargers.

It is important that you read and understand the information in this manual. The proper use of 
the 

SurePower Battery Pack

 is critical to the reliable operation of your ZOLL

 ®

 defibrillator.

Conventions

This manual uses the following conventions:

Warning!

Warning statements describe conditions or actions that can result in personal injury or 
death.

Caution

Caution statements describe conditions or actions that can result in damage to the 

SurePower 

Battery Pack

.

Note:

Notes contain additional information on using and maintaining the 

SurePower Battery 

Pack

.

The manual displays button and LED labels in 

bold

 face type.

Safety

The following sections provide important warnings, cautions, and notes regarding the ZOLL 

SurePower Battery Pack

.

Warnings

Locate ZOLL resuscitation devices where they cannot harm the patient or medical personnel 

if they fall -- accidental sliding or drop of the 

SurePower Battery Pack

 may cause physical 

injury. Damaged batteries may expose users to a safety hazard. Should the 

SurePower 

Battery Pack

 be dropped, inspect and test the unit prior to further use following the 

procedures described in this manual. 

Avoid prolonged exposure of the 

SurePower Battery Pack

 to direct sunlight to minimize 

heating and damage that may occur from exposure to ultraviolet light.

ZOLL rechargable defibrillator batteries have been tested against interference from radio 

frequency emissions typical of two-way radios used in emergency service/public safety 

activities. Refer to the Operator’s Guide for the ZOLL defibrillator to determine 

recommended operating distances from RF transmission equipment

Do not use ZOLL rechargable defibrillator batteries unless they have been inspected, tested, 

and recalibrated by appropriate personnel and you have been adequately trained in the 

proper use of these batteries.

Do not disassemble the 

SurePower Battery Pack

.

Charge the 

SurePower Battery Pack

 in a well-ventilated area.

You must perform periodic inspection, capacity testing, and functional testing of 

rechargable defibrillator batteries to ensure the availability of safe and reliable batteries.

If the battery in use on a ZOLL defibrillator indicates a fault condition or any operating state 

that can prevent the effective use of the defibrillator, replace that battery as soon as possible.

Do not place foreign objects into battery sockets.

Содержание SurePower

Страница 1: ...SurePower Battery Pack Guide 9650 0536 01 Rev B...

Страница 2: ...t ZOLL Medical Corporation to determine if additional product information updates are available ZOLL is a registered trademark of ZOLL Medical Corporation SurePower is a trademark of ZOLL Medical Corp...

Страница 3: ...fety hazard Should the SurePower Battery Pack be dropped inspect and test the unit prior to further use following the procedures described in this manual Avoid prolonged exposure of the SurePower Batt...

Страница 4: ...enance instructions If a problem occurs obtain service immediately Do not use alcohol or ketones MEK acetone etc to clean ZOLL rechargable defibrillator batteries Do not sterilize or autoclave ZOLL ba...

Страница 5: ...lows it to provide the following information The battery s run time or capacity The battery s calibration status Battery Faults Indications Figure 1 1 ZOLL SurePower Battery Pack Note ZOLL ships the S...

Страница 6: ...proximately 10 seconds to indicate the battery s estimated run time If the battery is fully charged the LED over each of the five runtime indicator icons illuminates As the battery discharges the numb...

Страница 7: ...ack Icon LED Indications This is the Recalibration icon This LED lights for approximately 10 seconds after you press and release the Display button if the battery needs to be calibrated If the Recalib...

Страница 8: ...onds While you hold down the Display button the runtime indicator displays the battery runtime When you release the Display button the runtime indicator displays the battery capacity for 2 seconds the...

Страница 9: ...ore the LED blinks The fault codes number from 1 to 29 To indicate fault code 1 the Fault LED blinks once To indicate fault code 2 the LED blinks twice and so on If the battery detects multiple batter...

Страница 10: ...6 5 Length x 5 7 Width x 4 4 Height cm Weight 1 7 lb 77 kg Voltage 10 8 V Capacity 5 8 Ah Operating Temperature 32 F to 122 F 0 C to 50 C Optimal Charging Temperature 59 F to 95 F 15 C to 35 C Short T...