7. Reassemble in reverse order. Place the washer into the nylon stem as shown in Figure
m
. Apply a drop of
locking fluid to the tip of the thread near the end of the screw. Push the screw up though the end of the stem
and through the washer.
8. Finally, slip the spring/plunger on the handle down into the stem. Rotate the stem until the slot in the stem
aligns with the rib in the handle and slips all the way into the handle, then tighten the screw.
9. Wait 24 hours before using the device to ensure that the locking resin obtains full strength. During this time,
the ResQPUMP should be hung by the strap or left resting on the handle with the suction cup pointing up.
7. TECHNICAL SPECIFICATIONS
Operating Specifications
Force gauge compression range
0 - 50 kg ± 15%
Force gauge decompression range
0 - 15 kg ± 15%
Metronome Function
Signal pitches
768 Hz (low) and 3070 Hz (high)
Sound level
≥65 dB at distance of 0.5 m for sound source
Rate
80 per minute
Temperature Range
Operation
-18º to 50º C (0º to 122º F)
Storage
-40º to 60º C (-40º to 140º F)
Dimensions
Suction cup
13.5 cm OD
Height
17.0 cm
Weight
0.58 kg
Materials
Suction cup and cushion pad
Silicone
Handle and support ring
Polyamide (nylon), glass fiber reinforced
Connection stem
Acetal polyoxymethylene
Support ring
Thermoplastic polyester elastomer
Metal parts
Stainless steel and brass
Battery
3.6 V primary lithium-thionyl chloride
Battery life:
Minimum of 20 service hours
Battery shelf life:
Approximately 10 years
8. ACCESSORIES AND REPLACEMENT COMPONENTS AVAILABLE
Reorder #
Description
12-0586-000
Suction Cup for ACD-CPR Device (replacement suction cup with support ring and
compression cushion)
12-0935-000
ResQCPR Carrying Case (holds one ResQPUMP and two ResQPODs)
12-2116-000
ManiKIT (permits many CPR manikins to be adapted for ACD-CPR)