12-Lead ECG Monitoring Setup
9650-000820-01 Rev. K
Propaq M Operator’s Guide
13-11
You can then view four 12-Lead snapshot pages by pressing
to step through them
sequentially. For example, the unit displays the following 12-Lead snapshot after the
Analysis
Page 2
:
Fault Conditions Affecting 12-Lead Interpretive Analysis
The Propaq M unit will not perform Interpretive Analysis if it detects one of the following fault
conditions when attempting to acquire 12-Lead data:
•
A pacer signal is present.
•
A lead fault is detected in the ECG cable.
•
An invalid cable is in use.
If the Propaq M unit detects one of these fault conditions,
Analysis Page 1
indicates that there is
No data available
for Interpretive Analysis and lists the fault condition; all measurements on
Analysis Page 2
appear as
N/A.
Once you have corrected the fault condition, press
to confirm the correction and perform
the 12-Lead Interpretive Analysis.