Xenon lamp
Light Source Housing
8
Zolix Instruments Co., Ltd.
http://www.zolix.com.cn E-mail:[email protected]
adjusting red knob 5, make the reflected light spot moving at 45 degree angle; By adjusting knob
3
、
4
、
5, make reflected light spot size and location coincide with light spot.
6) Parallel light adjustment. Firstly, you need to put a blank paper in about 250mm from the light
outlet, then by moving the blank paper(to the direction of the light outlet), you can observe the size
of light spot on the blank paper. if the size of light spot change obviously,you need to slide lens
cone until the size of light spot don’t change again.
7) If it is a multi outlet type light source in the room, according to the above steps before shooting the
main light path adjustment, to meet the use requirements by adjusting the optical path
corresponding to the side lamp side light path can be.
Note: the beam quality of the auxiliary light path may not be consistent with the main light path. When the actual
effect has been affected, the main optical path should be adjusted repeatedly so as to meet the use requirements.
3.3.3 Safety precautions
1) The output spectra of xenon lamps contain high levels of short wavelength ultraviolet radiation,
which can cause permanent damage to unprotected skin and eyes. Therefore, never look directly
into the output beam, any diffuse reflection, or any specular (mirror) reflection even for a short
period of time without suitable protective ultraviolet blocking filters or glasses.
2) Xenon lamps are always under pressures above atmosphere, so use extreme care when handling,
installing or mounting them to avoid any mechanical strains on the lamp. Protective glasses, heavy
gloves and a mask should always be worn when handling xenon lamps. Mechanical strain can
cause the lamp to shatter violently!
3) All xenon lamps, other than ozone free lamps, allow ozone gas to be generated while the lamp is
in operation. Always provide for proper ventilation to remove this ozone. High concentrations of
ozone are dangerous and can constitute a serious health hazard!
4) While the lamp is in operation, the instantaneous voltage will be more than 33,000 V and the
current will be larger than 25 A. Please let the lamp anode and the pole connection on trigger will
be contacted tightly.
5) In order to get the better light source stability, the lamp needs 30min preparing time.
6) The lamp in operation is very hot. Therefore, never touch the lamp envelops when it is in operation
or be turned off within 20 minutes.
7) In addition, please not cover the lamp housing with anything, because the Xenon lamp will
produce much heat when working.