VETSCAN UA USER’S MANUAL
|
2 7
7. Place strip on the Strip Tray and push the strip forward to the notch or “stop”. The end
of the strip should stop at the indent in the top cover of the UA analyzer, as with normal
sample runs. (Fig. 24)
This step should be completed before the countdown timer reaches
35.
The UA analyzer will time the reactions and scan when appropriate.
Fig. 24 Place strip onto Strip Tray
8. Results will appear onscreen when they are completed.
9. Compare results to values on UA Controls Package Insert.
10. Repeat steps 4 – 7. with a second UA test strip and the bottle of Control II. Results should
match the values on the Controls Package insert. If they do not match, check that strips
are not expired and that the strips have not been exposed to humidity (evidenced by
discoloration). Contact Zoetis Technical Support if the control results do not match expected
Control values after running them twice.
Urine Analyzer
Strip Type UA14
Species Dog
PID 1234
When beeping starts:
Quickly apply urine
Blot edge of strip
Place strip on holder.
Testing. . .
Cancel
35
35s