9. Results will appear onscreen one minute after start of the test (Fig. 15). The UA analyzer
will automatically print the results if Auto-Print is selected in the settings and the printer is
connected and powered on. Results may be manually printed via the Printer button at the
bottom left of the Results screen.
In general, urine dipsticks including those for the VETSCAN UA contain a number of different
parameters, some of which are not reliable or have not been well linked to disease processes in
veterinary medicine. Parameters which should be ignored include nitrites, leukocytes, ascorbic
acid, calcium and urobilinogen
1
. For improved precision, urine specific gravity is best assessed
with a refractometer and color is best assessed visually. The primary use of CRE on the UA14
strip is for calculating the urine protein: creatinine ratio, and CRE should not be evaluated on
its own.
Fig. 15 UA Results screen.
PID 1
Dog UA14
LEU
KET
NIT
URO
BIL
GLU
PRO
SG
pH
BLD
*ASC
MA
*Ca
CR
PRO/CR
-
-
Pos
Normal
-
-
-
1.045
5.0
-
+1
0|eu/uL
0mg/dL
0mg/dL
0mg/dL
<=15mg/dL
0mg/dL
25mg/dL
0.8mg/dL
<=4.0mg/dL
100mg/dL
<0.2
06:30:15PM
08/26/2017
Strip Type
Analyte names
Species
Patient ID
Sample Run
Date/Time
Numerical Results,
Units
Semi-quantitative Results
1
Sink CA and Weinstein NM. Routine Urinalysis: Chemical Analysis In: Practical Veterinary Urinalysis. Ames, IA: John Wiley & Sons Inc. 2012. pgs.
29-53.