VETSCAN UA USER’S MANUAL
|
1 7
5. Mix urine sample well by inversion. (Fig. 11).
6. Apply urine sample to strip. One method for application is to use a dropper pipette to drop
urine over entire strip, completely wetting each pad. Do not leave excess urine on strip
for more than 2 seconds (Fig. 12A). Alternatively, strip may be dipped into urine sample,
completely wetting each pad. Remove the strip after 2 seconds (Fig. 12B).
5X
Fig 12 Two Ways to Apply Urine to Strips (A) Drop sample on strip (B) Dip strip into sample.
Fig. 11 Mixing Sample
Fig. 12A
Fig. 12B