VETSCAN UA USER’S MANUAL
|
2 2
2. To clean the Strip Tray, first power on the UA analyzer. Remove the Strip Tray by grasping
the upper right side of the tray (Fig. 17A). Lift up and gently pull Strip Tray (but not the Tray
Carriage) in the direction shown (Fig. 17B). The Strip Tray should lift off easily.
Fig. 17A
Fig. 17B
3. Clean the Strip Tray using alcohol wipes or a lint free tissue with mild soap and water to
remove all residual urine (Fig. 18). Rinse with water to remove any cleaning substance which
may affect results. Never use corrosive cleaners, abrasive detergents or rough material that
may damage the surface. Cleaning the perimeter of the inner tray will prolong the useful life
of the system. Allow Strip Tray to completely dry prior to replacing it into the analyzer.
Fig. 18 Wipe Urine Residue from Strip Tray
Urine Analyzer