ZWave briefly
1.4. Handpiece
1
13
The handpiece (7) contains the pulse generator, a fan to dissipate heat
and the slot for the applicator head. It is connected to the control unit (1).
Note:
The pulse generator in the handpiece is an expendable part and has to be
replaced after a specific period of use as its functionality decreases over
time.
Zimmer MedizinSysteme guarantees unrestricted use of at least 2 million
pulses per pulse generator.
Wear of the pulse generator varies. Depending on performance and
frequency, sometimes far more than 2 million pulses can be delivered.
For more information on the need to replace the pulse generator, see
chapter 11.
To work with the handpiece on a client, it is essential that one of the
applicator heads is screwed tightly into the handpiece as far as possible.
The cable should not be stretched beyond its maximum length and must
be protected against pinching or any other mechanical damage.
To avoid heat accumulation in the handpiece, it is essential to ensure that
the air vents at the top and particularly on the base of the handpiece are
not blocked by the hand holding it or anything else.
Standby mode on device
and handpiece
The fan in the handpiece is started by depressing the footswitch and stops
automatically by reaching a certain temperature after the end of treatment.